Module 6
Low Current Measurement
6-38
Gate Oxide Capacitor
Chuck Guard
SMU1
SMU2
Chuck
10^13 ohms
10^13 ohms
Shielding Box (Ground)
Low Current Gate Oxide Meas.
Using Guarded Chuck
This is a simplified block diagram of the prober requirements for a guarded chuck connection. When
implemented properly, very fast low level sweeps are possible due to the elimination of stray
capacitance at the probes (wafer top side) as well as in the chuck (wafer bottom side).