Agilent 75000 SERIES B
Agilent E1364A 16-Channel
Form C Switch Module
User’s Manual
Copyright© Agilent Technologies, Inc., 2006
*E1364-90002*
E1364-90002
E0406
Manual Part Number: E1364-90002
Microfiche Part Number: E1364-99002
Printed: April 2006 Edition 2 Rev 2
Printed in Malaysia E0406
Vedere la pagina 0
1 2 3 4 5 6 ... 75 76

Sommario

Pagina 1 - *E1364-90002*

Agilent 75000 SERIES BAgilent E1364A 16-ChannelForm C Switch ModuleUser’s ManualCopyright© Agilent Technologies, Inc., 2006*E1364-90002*E1364-90002E0

Pagina 2

Notes8 Agilent E1364A 16-Channel Form C Switch User’s Manual

Pagina 3 - Contents

NotesAgilent E1364A 16-Channel Form C Switch User’s Manual 9

Pagina 4

Notes10 Agilent E1364A 16-Channel Form C Switch User’s Manual

Pagina 5

Chapter 1Getting Started with the Agilent E1364AUsing This ChapterThis chapter includes a Form C Switch description, addressing guidelines,and an exa

Pagina 6

Each channel is switched by opening or closing the appropriate channelrelay. Since the relays are latching, the relay remains in the last state during

Pagina 7

TypicalConfigurationThe Form C Switch accepts user inputs up to 250 V dc or 250 V ac peak(177 V ac RMS) at 1 A dc or ac RMS (non-inductive). Maximum r

Pagina 8 - Safety Symbols

You can address single channels (@ccnn); multiple channels(@ccnn,ccnn,...); sequential channels (@ccnn:ccnn); groups of sequentialchannels (@ccnn:ccnn

Pagina 9 - DECLARATION OF CONFORMITY

Example: Form CSwitch ChannelLists/RangesChannel Lists:CLOS (@100,112) Close channels 00 and 12 on card01OPEN (@203,210) Open channels 03 and 10 on ca

Pagina 10

16 Getting Started with the Agilent E1364A Chapter 1

Pagina 11

Chapter 2Configuring the Agilent E1364A Form CSwitchUsing This ChapterThis chapter shows how to make user connections to the Form C Switch andhow to

Pagina 13 - Instrument Definition

Caution STATIC-SENSITIVE DEVICE. Use anti-static procedures when removing,configuring, and installing a module. The Form C Switch is susceptible tosta

Pagina 14

Figure 2-2. Wiring the Form C Switch Terminal BlockChapter 2 Configuring the Agilent E1364A Form C Switch 19

Pagina 15 - Programming the Switch

Typical Switch ConfigurationsTypical Form C Switch configurations are:• General purpose relay• Relay control• Digital outputGeneral PurposeRelay Confi

Pagina 16 - Switch Channel

Example: RelayControl ConfigurationFigure 2-4 shows one way to control external relays using the +5V/NOsource. For this example, channel 15 is dedicat

Pagina 17 - Initial Operation

Digital OutputConfigurationBy connecting jumpers and installing pullup/pulldown resistors, you canconfigure the switch as a digital output device.Note

Pagina 18

Changing Form C Switch ComponentsAs required, you can:• Change the Logical Address switch setting• Change interrupt priority jumper positions• Replace

Pagina 19 - Chapter 2

Module Front Panel ConnectionsIf you choose not to use the supplied connector block, Figure 2-7 shows theE1364A front panel and the connection diagram

Pagina 20 - Connecting User Inputs

Chapter 3Using the Agilent E1364A Form C SwitchUsing This ChapterThis chapter provides examples to use the Form C Switch for switchingchannels and sc

Pagina 21

Switching ChannelsFor general purpose relay operation, you can connect or disconnect a loadby opening or closing specified channel relays. By adding j

Pagina 22 - Typical Switch Configurations

Example: ControllingMicrowave RelaysThis example shows one way that the Form C Switch can be used to controlan (external) microwave relay. Figure 3-2

Pagina 23 - Control Configuration

ContentsAgilent E1364A Switch Module Table of ContentsWarranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 24 - Configuration

Example: ControllingRF Switches/StepAttenuatorsFigure 3-3 shows one way to drive Agilent 8761 SPDT RF Switches orAgilent 33300 Series Programmable Ste

Pagina 25 - As required, you can:

Example: DigitalOutput ConfigurationFigure 3-4 shows channel 00 configured for digital output operation. Whenthe channel 00 relay is open (NC connecte

Pagina 26

Example: MatrixSwitchingThe Form C Switch can be configured as a 4 x 4 single-wire matrix toconnect any combination of up to four user sources (S0, S1

Pagina 27 - Form C Switch Commands

Scanning Channels• For the Form C Switch, scanning channels consists of closing aspecified set of channels, one channel at a time. You can scan anycom

Pagina 28 - Switching Channels

Scanning ChannelsCommentsChannel List Can Extend Across Boundaries. For multiple-moduleswitchbox instruments, the channels to be scanned can extend ac

Pagina 29 - Microwave Relays

Chapter 4Understanding the Agilent E1364A Form CSwitchUsing This ChapterThis chapter describes some techniques to scan Form C switch channels andto u

Pagina 30 - Attenuators

Selects the Trigger SourceTRIG:SOUR defines the triggersource to advance the scan. Defaultis TRIG:SOUR IMM. TRIG:SOUR?queries the current trigger sour

Pagina 31 - Output Configuration

Trigger Hold (TRIG:SOUR HOLD)TRIG:SOUR HOLD preventsexecution of triggers until triggersource is changed. Can use TRIGcommand to trigger a switchboxse

Pagina 32 - Comments

Example: ScanningWith External DevicesThis example uses the mainframe Trig Out port to synchronize the Form CSwitch channel closures to an external me

Pagina 33 - Scanning Channels

Example: ScanningUsing Trig Out andEvent In PortsThis example uses the mainframe Trig Out and Event In ports tosynchronize Form C switch channel closu

Pagina 34

Storing States . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Recalling States .

Pagina 35 - Scanning Channels Commands

Using the Scan Complete BitYou can use the Scan Complete bit (bit 8) in the Operation Status Registerof a switchbox to determine when a scanning cycle

Pagina 36

Special 488.2 CommandsThis section contains information about the response of the switchboxinstrument to three IEEE 488.2 common commands.Storing Stat

Pagina 37

SCPI Error ResponsesThe SYSTem:ERRor? query requests a value from the instrument’s errorregister. This register contains an integer in the range [-327

Pagina 38 - With External Devices

Chapter 5Agilent E1364A Form C Switch CommandReferenceUsing This ChapterThis chapter summarizes Standard Commands for ProgrammableInstruments (SCPI)

Pagina 39 - Event In Ports

CommandSeparatorA colon (:) always separates one command from the next lower levelcommand as shown below:ROUTe:SCAN:MODE?Colons separate the root comm

Pagina 40 - Using the Scan Complete Bit

Parameters Parameter Types. The following table contains explanations and examplesof the parameter types you might see later in this chapter.Table 5-1

Pagina 41 - Special 488.2 Commands

SCPI Command ReferenceThis section describes the Standard Commands for ProgrammableInstruments (SCPI) commands for the Form C Switch. Commands are lis

Pagina 42 - SCPI Error Responses

ARMThe ARM subsystem selects the number of scanning cycles (1 to 32767) foreach INITiate command.Subsystem Syntax ARM :COUNt <number> MIN | MAX

Pagina 43 - Reference

:COUNt? ARM:COUNt? [MIN|MAX] returns the current number of scanning cycles setby ARM:COUNt. The current number of scan cycles is returned when MINor M

Pagina 44 - Commands

DISPlayThe DISPlay subsystem monitors the channel state of a selected module (orcard) in a switchbox. This subsystem operates only with mainframes whi

Pagina 45 - ARM:COUN 1;:TRIG:SOUR EXT

IEEE 488.2 Common Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61Command Quick Reference . . .

Pagina 46 - SCPI Command Reference

MONitor[:STATe] DISPlay:MONitor[:STATe] <mode> turns the monitor mode on or off.ParametersParameter Name Parameter Type Range of Values<mode&

Pagina 47

INITiateThe INITiate command subsystem selects continuous scanning cycles andstarts the scanning cycle.Subsystem Syntax INITiate :CONTinuous <mode

Pagina 48 - Parameters

:CONTinuous? INITiate:CONTinuous? queries the scanning state. With continuousscanning enabled, the command returns 1. With continuous scanningdisabled

Pagina 49

OUTPutThe OUTPut command subsystem enables or disables the "Trig Out" port ofthe Agilent E1300B/E130lB Mainframes or on the Agilent E1405/E1

Pagina 50

[ROUTe:]The ROUTe command subsystem controls switching and scanningoperations for Form C Switch modules in a switchbox.Subsystem Syntax [ROUTe:] CLOS

Pagina 51 - INITiate

CLOSe? [ROUTe:] CLOSe? <channel_list> returns the current state of thechannel(s) queried. channel_list has the form (@ccnn) (see[ROUTe:]CLOSe fo

Pagina 52

OPEN? [ROUTe:]OPEN? <channel_list> returns the current state of the channel(s)queried. channel_list has the form (@ccnn) (see [ROUTe:]OPEN forde

Pagina 53

SCAN:MODE [ROUTe:]SCAN:MODE NONE | VOLT can be used for the Form C Switch.Both NONE and VOLT set the mode for "no measurements". Thiscommand

Pagina 54 - [ROUTe:]

:OPERation:ENABle STATus:OPERation:ENABle <number> enables the Operation StatusRegister to set a bit in the Status Register. For the Form C Swit

Pagina 55

• Maximum Error Numbers/Messages in the Error Queue: Thequeue holds a maximum of 30 error numbers/messages for eachswitchbox. If the queue overflows,

Pagina 56

Notes4 Agilent E1364A Switch Module Table of Contents

Pagina 57

:CPON SYSTem:CPON <number> | ALL sets the selected module (card) in aswitchbox to its power-on state.ParametersParameter Name Parameter Type Ran

Pagina 58

Example Advancing Scan Using TRIGger CommandThis example uses the TRIGger command to advance the scan of asingle-module switchbox from channel 00 thro

Pagina 59 - <number> numeric 1 - 99

• Using Bus Triggers: To trigger the switchbox withTRIGger:SOURce BUS selected, use the IEEE 488.2 commoncommand *TRG or the GPIB Group Execute Trigge

Pagina 60

IEEE 488.2 Common CommandsThe following table lists the IEEE 488.2 Common (*) Commands that applyto the Form C Switch module. The operation of some of

Pagina 61 - HOLD discrete Hold Triggering

Command Quick ReferenceSCPI Commands Quick ReferenceCommand DescriptionABORt Abort a scan in progress.ARM :COUNt <number>:COUNt? [MIN | MAX]Mult

Pagina 62

Appendix AAgilent E1364A Form C Switch SpecificationsMaximum Input Voltage:C to NC or NO: Any term. to chassis:250VDC 250VDC250VAC RMS 250VAC RMS354

Pagina 63 - IEEE 488.2 Common Commands

Relay LifeElectromechanical relays are subject to normal wear-out. Relay lifedepends on several factors. The effects of loading and switching frequ

Pagina 64 - Command Quick Reference

Appendix BAgilent E1364A Form C Switch RegistersRegister DefinitionsThe Agilent E1364A Form C Switch is a register-based device. See Figure B-1 for r

Pagina 65 - <7µV per chan

Addressing the RegistersTo read or write to specific registers, you must use the register address.Since the addresses for Agilent 75000 Series B or Se

Pagina 66 - Relay Life

Channel EnableRegisterA read of the Channel Enable register (base + 0816) always returns FFFF16,regardless of the channel states.Writing to the Regist

Pagina 67 - Channel Enable Register

CertificationAgilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. AgilentTechn

Pagina 68 - Reading the Registers

68 Agilent E1364A Form C Switch Registers Appendix B

Pagina 69 - Writing to the Registers

Appendix CAgilent E1364A Form C Switch ErrorMessagesTable C-l lists the error messages associated with the Form C Switchmodule programmed by SCPI. Se

Pagina 70

70 Agilent E1364A Form C Switch Error Messages Appendix C

Pagina 71 - Messages

IndexAgilent E1364A 16 Channel Form C Switch!*CLS 61*ESE 61*ESE? 61*ESR? 61*IDN? 61*OPC 61*OPC? 61*RCL 61*RST 61*SAV 61*SRE 61*SRE? 61*ST

Pagina 72

EEnable Scanning 34End of Relay Life 64:ERRor 56Error Messages 69 - 70Error Responses 40Event In Port 37ExampleControlling Microwave Relays 27C

Pagina 73

Relay Replacement Strategy 64Relays, Form C 12RF Switches 28[ROUT:]CLOS 25[ROUT:]CLOS? 25[ROUT:]OPEN 25[ROUT:]SCAN 25[ROUTe:] Subsystem 52[ROU

Pagina 74

74 Agilent E1364A 16 Channel Form C Switch Index

Pagina 75

Frame or chassis ground terminal—typicallyconnects to the equipment’s metal frame.Alternating current (AC).Direct current (DC). Indicates hazardous vo

Pagina 76

DECLARATION OF CONFORMITYAccording to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 Agilent E1364A 16-Channel Form C Switch User’s Manual 7Manufac