Calibration 3
U2761A Service Guide 29
Table 3 - 9 Configuration for AC amplitude flatness (+20 dB range)
verification
U2761A Measurement
Function Amplitude Frequency Nominal Error
Sine 1.750 Vrms 100.000 kHz 0.000 dB ±0.2 dB
Sine 1.750 Vrms 160.000 kHz 0.000 dB ±0.35 dB
Sine 1.750 Vrms 250.000 kHz 0.000 dB ±0.35 dB
Sine 1.750 Vrms 400.000 kHz 0.000 dB ±0.35 dB
Sine 1.750 Vrms 630.000 kHz 0.000 dB ±0.35 dB
Sine 1.750 Vrms 1.00000 MHz 0.000 dB ±0.35 dB
Sine 1.750 Vrms 1.60000 MHz 0.000 dB ±0.7 dB
Sine 1.750 Vrms 2.50000 MHz 0.000 dB ±0.7 dB
Sine 1.750 Vrms 4.00000 MHz 0.000 dB ±0.7 dB
Sine 1.750 Vrms 6.30000 MHz 0.000 dB ±0.7 dB
Sine 1.750 Vrms 10.0000 MHz 0.000 dB ±0.7 dB
Sine 1.750 Vrms 11.5000 MHz 0.000 dB ±0.7 dB
Sine 1.750 Vrms 13.2000 MHz 0.000 dB ±0.7 dB
Sine 1.750 Vrms 15.2000 MHz 0.000 dB ±0.7 dB
Sine 1.750 Vrms 17.4000 MHz 0.000 dB ±0.7 dB
Sine 1.750 Vrms 20.0000 MHz 0.000 dB ±0.7 dB