8 Chapter 2
Option Installation
Update Procedure
Writing System Correction Data of Backup Memory to Removable
HDD
The following is a description of the procedures to write the system correction data stored
in the backup memory to the mounted removable HDD.
Step 1. Turn on the E5070B/E5071B.
Step 2. Press .
Step 3. Press
Load Project... on the softkey menu.
Step 4. A dialog box appears to allow you to select a program to be loaded. Select
RestoreSysCorFile.vba from the D:\Agilent\Service folder, and then press
Open button.
Step 5. Press .
Step 6. RestoreSysCorFile dialog box appears (Figure 2-3). Click
OK.
Figure 2-3 RestoreSysCorFile Dialog Box
CAUTION Once the program has ended, restart the E5070B/E5071B to enable the system correction
data.