Agilent Technologies Z5623A Option H48 5 -11
Performance Verification
Setting Up Limit Testing
Step 6. The Scale window shown in Figure 5-6 will pop up. When it does, set
the following values:
•Scale Per Division = 1 dB
Reference Level = 0 dB
Reference Position = 9 Divisions
When these values have been entered, the Scale window will look like
Figure 5-7. Click the OK button.
Figure 5-6 The Scale Window With Default Values
Figure 5-7 The Scale Window With Desired Values For Insertion Loss
Vedere la pagina 78