Device Hi-Speed Tests 3
Notes on USB Electrical Compliance Testing 39
3 Place the Test Switch (S1) in the TEST position. Verify the yellow TEST
LED is lit. You should see the transmitted test packet on the
oscilloscope as below.
4 Click OK to close the Test Instructions dialog.
EL_6 Rise Time
EL_6 Fall Time
EL_2 EL_4 EL_5 Data Eye and Mask Test
After viewing the test results, click OK to close the Test Instructions
dialog.
Vedere la pagina 38