Agilent Technologies 34410A Guida Utente Pagina 168

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 190
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 167
Remarks
You can read the statistical values at any time.
This function returns the statistical peak-to-peak value for a collection of
readings (since statistics were enabled). This is different than the 2nd
Display peak-to-peak measurement, which measures high and low peaks
over a single measurement aperture.
The instrument clears the stored statistical data when statistics are
enabled, when the CALCulate:FUNCtion command is sent while
CALCulate:STATe is ON, when the power has been off, when the
CALCulate:AVERage:CLEar command is executed, after a Factory Reset
(*RST command), after an Instrument Preset (SYSTem:PRESet command),
or after a function change.
Vedere la pagina 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 189 190

Commenti su questo manuale

Nessun commento