Chapter 16 345
GPIB Command Reference
MANP
16. GPIB
Command Reference
( F - N )
MANP
Syntax MANP <value>
MANP?
Description When the manual sweep function is on, sets a measurement point.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
After the execution of the command, the sweep range specified with the
“MANR”(346page) command is canceled and the query response is (0,0).
Query response {value}<newline><^END>
Equivalent key No equivalent key is available on the front panel.
<value>
Description A measurement point number (number assigned in order from the
left end assuming that the left end measurement point is 1).
Range 1 to the number of measurement points
Preset value 1
Resolution 1
Vedere la pagina 344