12 Installation Note 83650-90007
Figure 12. W51 RF Cable Location
42. Remove the two hex screws from the RF deck.
43. Carefully pull the RF deck out of the instrument.
44. If the instrument has a rear panel Aux Output, disconnect W48 from the A28 amp/multiplier.
45. Rotate the RF deck 180 degrees (RF Output toward the rear) assuring that all cables remain
attached. Refer to Figure 13 to secure the RF deck in the service position.