Appendix A 335
Error Messages
Non-Recoverable Firmware Error
Use the following procedure to re-enter the three calibration factors
that were erased when RAM is cleared. Use the GENERATOR
ANALYZER screen keys (to the left of the cursor control knob) to access
the required screens.
1. Access the RF GENERATOR screen and select DC FM Zero (under
the FM Coupling field).
2. Disconnect any cables from the ANT IN or RF IN/OUT connectors.
3. Access the RF ANALYZER screen and select Zero under the TX Pwr
Zero field.
4. Access the AF ANALYZER screen and select Zero under the
DC Current field.