The Agilent 4395A/96B can perform direct
impedance measurements with the addition
of Option 439xx-010 and the Agilent 43961A
RF impedance test kit. Covering from 100
kHz, impedance parameters |Z|, θ, C, L,
Q, D and more are measured and displayed
directly on the color TFT LCD display. For
the Agilent 4395A, the Option 4395A-001
DC source is available to apply to the
device, up to ±40 V maximum.
Direct display of impedance data
The impedance measurement capability
provides powerful display choices, allow-
ing direct parameter versus frequency
display. Measure and display capacitance,
inductance, and impedance magnitude and
phase, and more, as functions of frequency.
In addition, a lumped equivalent circuit
function aids component modeling and
simulation.
Many applicable fixtures and
powerful error compensation
Many 7-mm connector type test fixtures
can be used for the Agilent combination
analyzer with Option 439xx-010 and the
43961A impedance test kit because the
43961A employs the 7-mm connector as
the measurement port. So, you can select
an appropriate fixture among them. The
combination analyzer with Option 439xx-
010 has the powerful OPEN/SHORT/LOAD
compensation, and port extension compen-
sation, which eliminate any errors intro-
duced by the test fixture and allow accurate
direct impedance measurements.
Accurate Direct Impedance Measurement
Connecting the Agilent 43961A to the Agilent 4395A/96B
Two independent
measurement
channels and
flexibility to
impedance
analysis
Equivalent component analysis
(resonator)
Agilent 4395A/96B
Agilent 43961A
Test fixture
N-N cable
RF in
DC source
input
Output
R
A
R
A
BRF out
Vedere la pagina 9