Definition of Terms Specifications
Agilent 81600B Tunable Laser Source Family, Fourth Edition 45
Relative wavelength accuracy (continuous
sweep mode)
When measuring the differences between the actual and “Logged
wavelength” in “Continuous sweep mode”, the dynamic wavelength
accuracy is ± half the span between the maximum and the minimum value
of all differences.
Figure 14 Relative wavelength accuracy (continuous sweep mode).
Conditions: As specified. No mode-hop.
Relative wavelength accuracy (stepped
mode)
When randomly changing the wavelength and measuring the differences
between the displayed and the actual wavelength, the relative wavelength
accuracy is ± half the span between the maximum and the minimum value
of all differences.
Figure 15 Relative wavelength accuracy.