Chapter 5 73
Mode Parameters
Test Set Parameters
Mode Parameters
Figure 5-1 Front panel input connectors
To select the connector to be used as the input port, and the connector to be used as the output port, use
the
Input/Output hardkey on the front panel, or use the following SCPI remote control commands:
[:SENSe]:FEED:RF:PORT[:INPut] RF|RFIO1|RFIO2 (to select the input port);
[:SENSe]:FEED:RF:PORT:OUTPut RFOut|RFIO1|RFIO2 (to select the output port).
For more information about the SCPI commands refer to the Sequence Analyzer online help (or in pdf
format, the Users and Programmers Reference).