Agilent Technologies N2782B Manuale di Servizio Pagina 10

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 46
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 9
10 N2780/81/82/83 User’s and Service Guide
WARNING
If the waveform measuring instrument being connected to the output
terminal (BNC) on this device is equipped with any other
measurement terminals, take the following precautions to ensure that
the other instrument does not form a bridge between the probe and
any hazardous live part of a part.
1 Isolate the terminal to which the probe is connected from other
terminals on the measuring instrument using basic insulation
conforming to the measurement category
, working voltage, and
pollution degree
requirements of the circuit being tested.
2 If basic insulation
requirements cannot be met between the terminal
to which this device is connected and other terminals of the
measuring instrument, make sure that the voltage input to the
measurement terminal does not exceed the Separated Extra- Low
Voltage Earthed (SELV- E).
3 Read and observe all warnings and precautions relating to electrical
safety for the measuring instrument being connected to the probe.
Refer to the following standards regarding the meanings of underlined
terms.
IEC 61010- 1
IEC 61010- 031
IEC 61010- 2- 032
medium_standard.book Page 10 Friday, September 10, 2010 2:45 PM
Vedere la pagina 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 45 46

Commenti su questo manuale

Nessun commento