Agilent Technologies 1100 Series Manuale di Servizio Pagina 114

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 318
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 113
114 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Replacing Capillaries on a Standard Flow Cell
When required If the capillary is blocked
Tools required Two 1/4 inch wrenches for capillary connections
Wrench 4 mm for capillary connections
Screwdriver Pozi 1 PT3
Parts required For parts see “Standard Flow Cell” on page 186
Preparations for this procedure:
Turn off the flow.
Remove the front cover.
Remove the flow cell, see “Exchanging a Flow Cell” on
page 105.
1 Identify the inlet and outlet capillaries.
Outlet capillary
Inlet
capillary
Vedere la pagina 113
1 2 ... 109 110 111 112 113 114 115 116 117 118 119 ... 317 318

Commenti su questo manuale

Nessun commento