Chapter 1 1-47
Alphabetical Command Reference
CSWI
CSWI
Syntax
CSWI<ON|OFF>; or CSWI?;
NOTE These commands only apply to ES models.
Description
See also “TSSWI” on page 1-267.
Front Panel Equivalents
Command Description Range Query Response
CSWION Selects test set continuous switching.
When continuous switching is on, the
analyzer measures all four S-parameters
each time before displaying the data for a
full 2-port cal measurement.
N/A N/A
CSWIOFF Selects test set hold mode. The analyzer
measures all four S-parameters once and
then measures the desired parameter
continuously. This is known as a fast
2-port cal measurement and it is less
accurate than a full 2-port calibrated
measurement.
N/A N/A
CSWI? Queries whether continuous switching is
on or off.
N/A
<0|1><
L
F
>
Command Hardkey Softkey
CSWION
CSWIOFF
Cal
TESTSET SW CONTINUOUS
Cal
TESTSET SW HOLD