Get Assistance, if You Need It 6
Agilent E6474A User’s Guide 107
receiver requires a 30-minute warm-up period to meet the
guaranteed specifications.
2 Depending on your system type, set the frequency/power
levels using “Signal Generator Uplink Settings" on page 108.
3 Verify that the correct values are set on the signal generator.
NOTE
There is a three-minute delay after power-up or re-configuration, before
the internal temperature sensor is activated and measurements are
nominally correct.
Figure 50 System measurement verification setup using a signal generator
NOTE
Do not connect the signal generator at this stage.