安捷倫科技高頻元件量測研討會
Feb.23, 2006
Agenda
Overview of PNA FW Rev. 6.0
Cal Registers
TRL Calibration with 4-port 20 GHz PNA-L
Unknown Thru Calibration
Offset Load Cal
Fixture And Probe Techniques
Auto port extensions
Embedding/de-embedding
Frequency Converter Applications – SMC
Measuring fixtures/probes
Q&A
Page 10
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Old Calibration Paradigm
All calibrations were saved in a “calibration set”
“.cst” file pointed to data in a calibration set, but did NOT contain
calibration data itself
“.cal” files contained calibration data but NOT instrument
parameters
Saves instrument state using current state name
(does not automatically select a .cst file if current state is .sta)
Opens a file dialog box to save instrument state with
a new file name (choose .sta or .cst)
Saves instrument state with an auto-incremented name that
won’t overwrite existing states (e.g. “at004.cst”)
Old file save dialog