4Service
80 U2000 Series Operating and Service Guide
Reassembly Procedure
Tools required for reassembly:
Table 4-13 Attenuator Reassembly Procedure
Tools Purpose Torque value
Loctite
Threadlocker 242
To secure the connection
between the attenuator and the
sensor’s connector
N/A
¾” torque wrench To tighten the attenuator 12 lbs.in
1 Apply one drop of loctite on
the first, second, and third
threads of the connector.
2 Tighten the connector using the
torque wrench.