5. Set up the phase noise measurement system.
a) From the Define menu, choose Measurement;
then define the measurement per the parame-
ter data provided in Table 2-4. When all of the
forms have been defined, choose the Close
button.
6. Select View Meter (Tool Bar icon).
a) This internal meter provides quadrature status;
the black bar is not present when quadrature
is set.
7. Start New Measurement.
a) Select New Measurement from the Measure
pulldown menu.