Agilent Technologies E5500 Series Guida Utente Pagina 147

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 232
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 146
Scanner Maintenance and Calibration 8
Agilent 5500 SPM User’s Guide 147
X Sensitivity
Using the Cross-section tool, measure the length of a set of features
across the scan (Figure 100).
Figure 100 Cross-section of several features to check Xsensitivity
If the measured dimension does not match the actual, then adjust the X
Sensitivity term according to the following equation:
After adjustment the measured size should be within 5 % of the actual
size.
Y Calibration
The methods used for calibrating the scanners Y axis are similar to
those used for the X dimension. The exception is that the scan range will
be set to ½ of the scanners full range. The Y axis is generally used as
NewSensitivity
CurrentSensitivity Kno wnSize
MeasuredSize
-------------------------------------------------------------------------------------
=
Vedere la pagina 146
1 2 ... 142 143 144 145 146 147 148 149 150 151 152 ... 231 232

Commenti su questo manuale

Nessun commento