Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Arming Second Analyzer
Arming Second Analyzer
If you are using both analyzers on the same measurement module, you
can have one analyzer arm the other analyzer. The functionality of this
feature is handled in two different ways, depending on the analyzer
model.
167xx - Arming Second Analyzer
If you are using a 16715 and newer analyzer, the arming of one analyzer
by another is handled by the trigger function "Wait for second analyzer
to trigger".
165xx - Arming Second Analyzer
If you are using a 165xx analyzer, the arming of one analyzer by
another is handled by the "Machine Arming Tree" found under Arming
Control in the logic analyzer's Trigger tab - Settings tab.
The order in which the analyzers appear in the Arming Tree
determines the order they receive the arming signal. The arm signal
does not automatically start the next analyzer. The arm signal simply
tells the next analyzer that it can start evaluating its own trigger
specification and run when its trigger conditions are satisfied.
If a trigger specification in an analyzer has multiple sequence levels,