178 Chapter 6
BERT Testing
Measuring RF Loopback BER with Option 300
Figure 6-13
10. Press
Trigger to start the measurement:
After the search is complete, Pass or Fail is displayed on the lower left corner of the
screen, when either of the following occurs:
The result equals or is less than the target percent within the high to low amplitude
range.
The high or low amplitude level is passed at the target% BER/RBER.
Refer to Figure 6-14.