Repairing the Agilent
Model V24CT/V26CT
Rear Housing Removal Procedures
Disassembling the Rear Housing consists of:
Removing power to the Model V24CT/V26CT by pressing the power switch to OFF and
disconnecting the AC power cord.
Removing Lead Acid Batteries (see 9-5).
Removing Parameter Module Rack (see 9-6).
Removing the Front Housing Assembly (see 9-2).
Removing Battery Contact Board (see 9-7).
Removing the System Board (see 9-8).
Removing Power Supply Assembly (see 9-10).
Removing Parameter Module Rack Connector (see 9-13).
5HPRYLQJ%DWWHU\&RQWDFW%RDUG
Remove power to the Model V24CT/V26CT by pressing the power switch to OFF and
disconnecting the AC power cord.
Remove batteries. See 9-5.
Remove Front Housing Assembly. See 9-2.
Remove battery contact board power cable connector.
Remove four screws securing battery contact board to power supply assembly
Remove battery contact board.
Contact Board
Power Cable
Connector