114 Chapter 3
Basic Digital Operation
Using Waveform Clipping
Figure 3-16 Reduction of Peak-to-Average Power
Configuring Circular Clipping
This procedure shows you how to configure circular clipping. The circular setting clips the composite
I/Q data (I and Q data are clipped equally). For more information about circular clipping, refer to
“How Clipping Reduces Peak- to- Average Power” on page 111.
1. Press
Preset > Mode > Custom > Arb Waveform Generator > Digital Modulation Off On to On. This generates a
custom arbitrary waveform for use in this procedure. You can also use a previously stored or
downloaded waveform.