Calibrating the Infiniium Oscilloscope and Probe 14
U7232A DisplayPort Compliance Testing Notes 133
Figure 50 Oscilloscope Calibration Window
5 Follow the on- screen instructions:
a You will be prompted to disconnect everything from all the inputs:
click the OK button.
b Then you will be prompted to connect the calibration cable with
SMA adapters between the Aux Out and a specified input. Install the
SMA adapter by pressing it on input BNC, and hand tightening the
outer ring turning right. Click the OK button after connecting the
cable as prompted.
c During the calibration of channel 1, you will be prompted a
Calibration Options screen, as shown in
Figure 51 below.