To define RF4 I|O as an input and RF3 I|O as an output, use the following steps.
Define RF4 I|O as an input, using this menu setting or its equivalent command:
Input/Output > More > RF Output & Test Set Config > Half Duplex Config >
RF Input Port > RF4 I|O
Command:
:HDUPlex:PORT:INPut RFIO4
Define RF3 I|O as an output, using this menu setting or its equivalent command:
Input/Output > More > RF Output & Test Set Config > Half Duplex Config > RF
Output Port > RF3 I|O
Command:
[:SENSe]:HDUPlex:PORT:OUTPut RFIO3
NOTE This does not complete the port setup; two more settings are required!
Set up the RF Input using this menu setting or its equivalent command:
Input/Output > RF Input > RF Input Port > RF Input
Command:
:FEED:RF:PORT RFIN
Set up the RF Output using this menu setting or its equivalent command:
Input/Output > More > RF Output & Test Set Config > RF Output Port > RF Output
Command:
:FEED:RF:PORT:OUTP RFO