The individual frequencies are returned by the MEAS:JTOL:COMP(*):PAR:FLIS?
query.
MEAS:JTOL:COMP(*):PARameter:BTESt[?]
IAgilentN490xJComplianceMeasurement.BERTestMode
MEAS:JTOLerance:COMP(*):PARameter:BTESt CLEVel | ABSolute
MEAS:JTOL:COMP(*):PAR:BTESt?
The command sets the verification method to either confidence level (CLEVel) or
numbers of received bits/errors (ABSolute). In both cases, additional parameters
must be specified.
The query returns the present setting, CLEV or ABS.
MEAS:JTOL:COMP(*):PARameter:BTESt:TBER[?]
IAgilentN490xJComplianceMeasurement.TargetBER
MEAS:JTOLerance:COMP(*):PARameter:BTESt:TBER <NR3>
MEAS:JTOL:COMP(*):PAR:BTESt:TBER?
This command specifies the target bit error ratio, for example 1.0e-12.
The query returns the present value.
MEAS:JTOL:COMP(*):PARameter:BTESt:CLEVel[?]
IAgilentN490xJComplianceMeasurement.TargetBERConfidence
MEAS:JTOLerance:COMP(*):PARameter:BTESt:CLEVel <NR3>
MEAS:JTOL:COMP(*):PAR:BTESt:CLEV?
This command is needed if the verification method is set to CLEVel. It sets the
confidence level, for example 0.97. The setting takes effect when the confidence
level CLEVel is enabled by means of the MEAS:JTOL:COMP(*):PAR:BTESt
command.
The query returns the present setting.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 369
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description