4
Agilent N4916B Applications
Transmitter (TX) Receiver (RX)Channel
ChannelTX latch
TX PLL
RX latch
CDR
RX PLL
Clock Clock
Reference
Clock
No TX de-emphasis Closed eye at receiver input
With TX de-emphasis Open eye at receiver input
Signal before channel Signal after channel
Figure 4. De‑emphasis helps to compensate for the signal degradations caused by the channel.
Also some of the signal degradations caused by cables, fixtures or test boards in the test setup
can be compensated for using de‑emphasis.
De-embedding signal degradations
The N4916B can also be used to compensate for some of the signal degradations caused
by the test set up; e.g. cables, fixtures, test boards. This is helpful to optimize the jitter
budget needed for accurate receiver tolerance characterization.