2 - Verification and Performance Tests
18
c. Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to
amps and record this value (Iout). The readings should be within the limits specified in the performance test
record card for the appropriate model under Current Programming @ 0 Amps.
d. Set the current range readback to High and program the output current to 20mA. Repeat step C to get the Iout.
Record the current reading on the front panel display. The reading should be within the limits specified in the
performance test record card for the appropriate model under Current Readback Accuracy (20mA) Iout.
e. Program the output current to the full-scale value in Table 2-2.
f. Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to
amps and record this value (Iout). Also, record the current reading that appears on the front panel display. The
readings should be within the limits specified in the performance test record card for the appropriate model under
Current Programming and Readback @ Full Scale.
Current Sink (-CC) Operation
This test verifies current sink operation and readback.
a. Turn off the supply and connect the output as shown in Figure 2-1a, except connect a dc power supply in place
of the electronic load as indicated. Set the DMM to operate in voltage mode.
b. Set the external power supply to 5 V and the current to the full scale current rating of the supply under test as in
Table 2-2.
c. Turn on the supply under test and program the output voltage to zero and the current to full scale as in Table 2-
2. The current on the UUT display should be negative and at least 60% of the current rating.
d. Divide the voltage drop across the current monitoring resistor by its resistance to obtain the current sink value
in amps and subtract this from the current reading on the display. The difference between the readings should be
within the limits specified in the performance test record card under Current Sink Readback.
Low Range Current Readback Accuracy
This test verifies the readback accuracy of the 20 milliampere current range.
a. Turn off the supply and connect the output as shown in Figure 2-1b. Set the DMM to operate in current mode.
b. Turn on the supply under test and set the current range readback to Low. Program the output voltage to zero
and the current to the full scale value in Table 2-2. The current on the UUT display should be approximately 0
mA.
c. Record the current reading on the DMM and the reading on the front panel display. The difference between the
two readings should be within the limits specified in the performance test record card under 20mA Range Current
Readback Accuracy @ 0A.
d. Program the output voltage to 8V and record the current reading on the DMM and the reading on the front
panel display. If the meter indicates overrange, lower the 8 volts slightly. The difference between the readings
should be within the limits specified in the performance test record card for the appropriate model under 20mA
Range Current Readback Accuracy @ +20mA
e. Turn off the supply and connect the output and an external supply as shown in Figure 2-1c. Set the DMM to
operate in current mode.
f. Turn on the external supply and program it to 8V and 1 amp. Then program the supply under test to zero volts
and 1 amp , except the 6614C is programmed to 0.5 amp . If the meter indicates overrange, lower the voltage of the
external supply slightly. The UUT display should read approximately 20 mA.
g. Record the current reading on the DMM and the reading on the front panel display. The difference between the
two readings should be within the limits specified in the performance test record card under 20mA Range Current
Readback Accuracy @ 20 mA.
Vedere la pagina 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 58 59

Commenti su questo manuale

Nessun commento

ASCOM OFFICE 130PRO manuals

Owner’s manuals and user’s guides for Telephones ASCOM OFFICE 130PRO.
We providing 3 pdf manuals ASCOM OFFICE 130PRO for download free by document types: Operations Instructions, User Manual, User's Guide






More products and manuals for Telephones ASCOM

Models Document Type
OFFICE 25 Operations Instructions   ASCOM OFFICE 25 Operating instructions, 52 pages
Eurit 33 Plus User Manual   Ascom Eurit 33 / 33 plus, 100 pages
Eurit 20 User Manual   ascom Eurit 30, 35 pages
OFFICE 10 User Manual Ascotel Office Terminals by ascom, 68 pages
Eurit 33 Plus Operations Instructions   ASCOM Eurit 33 Plus Operating instructions [en] , 100 pages
D81 - User Manual   ASCOM D81 - User manual [en] [es] , 120 pages
21 User Manual   ASCOM 21 User manual, 57 pages
I62 - User Manual   ASCOM I62 - User manual, 22 pages
i75 VoWiFi User Manual   Application Note for Configuring the Ascom wireless i75 VoWiFi, 24 pages
TD 92582GB User Manual   ASCOM TD 92582GB User manual, 48 pages
i75 VoWiFi User Manual   Application Note for Configuring the Ascom wireless i75 VoWiFi [en] , 26 pages
TD 92582GB User Manual   ASCOM TD 92582GB User manual, 23 pages
21 User Manual   ASCOM 21 User manual, 108 pages
Eurit 20 User Manual   ascom Eurit 20, 26 pages
OfficeT User Manual   Application Note for Configuring the Ascom Wireless IP, 22 pages
Eurit 33 Plus User Manual   ASCOM Eurit 33 Plus User manual, 105 pages
TD 92582GB User Manual   ASCOM TD 92582GB User manual, 32 pages
I62 - User Manual   ASCOM I62 - User manual, 78 pages
d41 DECT User Manual   Application Note for Configuring the Ascom Wireless IP, 27 pages
Avena 122 plus User's Guide    ASCOM Avena 122 plus User guide, 74 pages