28 Multimode Source for 6100 Series Single Quad LC/MS User Guide
2 Installation
Step 2. Install the HV control PCA and cables
Figure 16 Installing the multimode source (left) and connecting all connectors.
14 Put back the side, top, front and CDS cover.
15 Plug the system power cord back on and turn the front switch on.
The pump down process will start.
16 Start the ChemStation program.
17 Click Method and Run Control View > MSD > Spray Chamber and verify
that the source is an MM Source.
18 Go to MSD Tune window and start the Auto tune with MM source.
19 Verify that the multimode source is working correctly.