100 Agilent FrameScope 350 User’s Manual
4 Fiber Testing
Calibrating the Tester
6 Choose Single Ended, and press Next. The Fiber Detection screen
displays.
7 When the message below the diagram reads, “Ready to Calibrate,
press OK. The Calibration window opens.
8 When the window reads, “Calibration Successful,” press OK in the
window. The Fiber Main Menu displays.