Cable Tester Reference 9
Calibration
Agilent FrameScope 350 User’s Manual 217
To run a self test on the DualRemote 350:
1 Insert the Self-Test probe in the SmartProbe port of the DualRemote
350.
2 Press the OK key. The Pass/Fail LED on the front panel of the
DualRemote 350 indicates the outcome of the self test.
If the self test fails, the DualRemote 350 requires a new baseline
calibration or repair. Contact Agilent Technical Support.