How to switch batteries during calibration
If, during calibration, you need to change the battery supply
for the timing module, follow this procedure:
1 Connect the AC adaptor power supply to the booster
cable charging port.
2 Ensure the AC power is on.
3 Carefully remove the battery from the cable.
4 Replace the battery with a fully charge unit.
5 Remove the AC adaptor power supply.
Cable connections for indoor hardware
Once the timing module has been calibrated, you can
connect the hardware for indoor measurement. Refer to
“How to add and configure an indoor map" on page 98 for
more information on configuring the E6474A software.
Refer to Figure 25 on page 52 for cable connection
information. This diagram does not show the backpack and
battery pouches.
Do not disconnect power to the timing module after the calibration period.
This may cause the lose of the timing synchronization 1PP2S signal.