3- 18 Agilent B1500A Configuration and Connection Guide, Edition 5
Connection Guide for Wafer Prober
SCUU and ASU
NOTE About ASU output cables
To perform capacitance measurement accurately, the cable length between DUT and the
ASU output must be as short possible.
NOTE For the installation of the ASU and the connection to the DUT interface (probe card,
manipulators, and so on), contact your favorite prober vender. The prober vender will have
the solutions. Dimensions of the ASU are 132 mm (W) 88.5 mm (H) 50 mm (D)
excluding the connectors.