Sommario

Pagina 1 - Oscilloscopes

 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide

Pagina 2

10 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide :CALibrate:LABel 214:CALibrate:OUTPut 215:CALibrate:PROTected 216:CALibra

Pagina 3 - In This Book

100 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :SBUS<n>:UART:SOURce:TX <source> (

Pagina 4

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 101 Table 28 General :SEARch Commands SummaryComma

Pagina 5

102 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :SEARch:GLITch:RANGe <less_than_time>[su

Pagina 6

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 103 :SEARch:TRANsition:SOURce <source> (see

Pagina 7 - Contents

104 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :SEARch:SERial:I2S:PATTern:DATA <string>

Pagina 8

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 105 :SEARch:SERial:IIC:PATTern:DATA2 <value>

Pagina 9

106 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference Table 37 :SEARch:SERial:SPI Commands SummaryCo

Pagina 10 - 13 :DISPlay Commands

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 107 Table 39 :SYSTem Commands SummaryCommand Query

Pagina 11 - 14 :EXTernal Trigger Commands

108 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :TIMebase:REFerence {LEFT | CENTer | RIGHt} (s

Pagina 12 - 19 :MEASure Commands

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 109 :TRIGger:MODE <mode> (see page 648):TRIG

Pagina 13 - 20 :MTESt Commands

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 11 :DISPlay:DATA 258:DISPlay:LABel 259:DISPlay:LABList 260:DISPlay:PERSisten

Pagina 14 - 23 :SAVE Commands

110 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference Table 43 :TRIGger[:EDGE] Commands SummaryComma

Pagina 15 - 24 :SBUS<n> Commands

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 111 Table 44 :TRIGger:GLITch Commands SummaryComma

Pagina 16

112 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :TRIGger:GLITch:RANGe <less_than_time>[s

Pagina 17 - 25 :SEARch Commands

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 113 :TRIGger:PATTern:QUALifier <qualifier> (

Pagina 18 - 26 :SYSTem Commands

114 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :TRIGger:SHOLd:SOURce:DATA <source> (see

Pagina 19 - 28 :TRIGger Commands

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 115 Table 49 :TRIGger:TV Commands SummaryCommand Q

Pagina 20

116 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference Table 51 :WAVeform Commands SummaryCommand Que

Pagina 21 - 31 :WMEMory<r> Commands

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 117 n/a :WAVeform:PREamble? (see page 724)<prea

Pagina 22

118 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference n/a :WAVeform:XINCrement? (see page 737)<re

Pagina 23 - 35 Synchronizing Acquisitions

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 119 :WGEN:OUTPut:LOAD <impedance> (see page

Pagina 24 - 37 Programming Examples

12 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide :HARDcopy:PALette 301:HARDcopy:PRINter:LIST 302:HARDcopy:STARt 30317 :LIS

Pagina 25 - What's New

120 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :WMEMory<r>:SKEW <skew> (see page

Pagina 26 - 1 What's New

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 121 Syntax Elements• "Number Format" on

Pagina 27 - Version 1.00 at Introduction

122 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference For example, <A> ::= <B> indicates

Pagina 28

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 123 #800001000<1000 bytes of data> <NL>

Pagina 29

124 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference

Pagina 30

125 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5Common (*) CommandsCommands defined by IEEE 488.2 standard that are co

Pagina 31 - Discontinued

126 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands n/a *OPT? (see page 137) <return_value> ::= 0

Pagina 32

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 127 Introduction toCommon (*)CommandsThe common command

Pagina 33 - Setting Up

128 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands selected and a common command is received by the in

Pagina 34 - 2 Setting Up

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 129 *CLS (Clear Status) (see page 860)Command Syntax*CL

Pagina 35 - Using the LAN Interface

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 13 :MEASure:RESults 356:MEASure:RISetime 359:MEASure:SDEViation 360:MEASure:

Pagina 36 - Using the GPIB Interface

130 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands *ESE (Standard Event Status Enable) (see page 860)C

Pagina 37 - Setting Up 2

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 131 The *ESE? query returns the current contents of the

Pagina 38

132 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands *ESR (Standard Event Status Register) (see page 860

Pagina 39 - Click Test Connection

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 133 See Also • "Introduction to Common (*) Command

Pagina 40

134 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands *IDN (Identification Number) (see page 860)Query Sy

Pagina 41

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 135 *LRN (Learn Device Setup) (see page 860)Query Synta

Pagina 42

136 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands *OPC (Operation Complete) (see page 860)Command Syn

Pagina 43 - Getting Started

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 137 *OPT (Option Identification) (see page 860)Query Sy

Pagina 44 - Capturing Data

138 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands • "*IDN (Identification Number)" on page

Pagina 45 - Analyzing Captured Data

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 139 *RCL (Recall) (see page 860)Command Syntax*RCL <

Pagina 46 - Programming the Oscilloscope

14 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide :MTESt:RMODe:FACTion:PRINt 406:MTESt:RMODe:FACTion:SAVE 407:MTESt:RMODe:F

Pagina 47

140 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands *RST (Reset) (see page 860)Command Syntax*RSTThe *R

Pagina 48 - 3 Getting Started

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 141 Digital Channel Menu (MSO models only)Channel 0 - 1

Pagina 49 - Getting Started 3

142 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands See Also • "Introduction to Common (*) Command

Pagina 50 - Ensure New Data is Collected

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 143 *SAV (Save) (see page 860)Command Syntax*SAV <va

Pagina 51

144 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands *SRE (Service Request Enable) (see page 860)Command

Pagina 52

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 145 Query Syntax *SRE?The *SRE? query returns the curre

Pagina 53

146 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands *STB (Read Status Byte) (see page 860)Query Syntax*

Pagina 54 - Checking Instrument Status

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 147 See Also • "Introduction to Common (*) Command

Pagina 55 - Telnet Sockets

148 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands *TRG (Trigger) (see page 860)Command Syntax*TRGThe

Pagina 56

Common (*) Commands 5Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 149 *TST (Self Test) (see page 860)Query Syntax*TST?The

Pagina 57 - Commands Quick Reference

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 15 :SAVE:WMEMory:SOURce 450:SAVE:WMEMory[:STARt] 45124 :SBUS<n> Comman

Pagina 58 - Command Summary

150 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide5 Common (*) Commands *WAI (Wait To Continue) (see page 860)Command Synta

Pagina 59

151 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6Root (:) CommandsControl many of the basic functions of the oscillosco

Pagina 60

152 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands :BLANk [<source>] (see page 162)n/a <source&

Pagina 61

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 153 :OVLenable <mask> (see page 175):OVLenable? (se

Pagina 62

154 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands Introduction toRoot (:)CommandsRoot level commands co

Pagina 63

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 155 :ACTivity (see page 860)Command Syntax:ACTivityThe :A

Pagina 64

156 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands :AER (Arm Event Register) (see page 860)Query Syntax:

Pagina 65

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 157 :AUToscale (see page 860)Command Syntax:AUToscale:AUT

Pagina 66

158 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands For further information on :AUToscale, see the User&a

Pagina 67

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 159 :AUToscale:AMODE (see page 860)Command Syntax:AUTosca

Pagina 68

16 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide :SBUS<n>:IIC[:SOURce]:DATA 497:SBUS<n>:IIC:TRIGger:PATTern:AD

Pagina 69

160 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands :AUToscale:CHANnels (see page 860)Command Syntax:AUTo

Pagina 70

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 161 :AUToscale:FDEBug (see page 860)Command Syntax:AUTosc

Pagina 71

162 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands :BLANk (see page 860)Command Syntax:BLANk [<source

Pagina 72

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 163 :DIGitize (see page 860)Command Syntax:DIGitize [<

Pagina 73

164 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands Example Code ' Capture an acquisition using :DIG

Pagina 74

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 165 :MTEenable (Mask Test Event Enable Register) (see pag

Pagina 75

166 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands • ":CHANnel<n>:PROTection" on page 2

Pagina 76

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 167 :MTERegister[:EVENt] (Mask Test Event Event Register)

Pagina 77

168 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands • "*STB (Read Status Byte)" on page 146• &q

Pagina 78

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 169 :OPEE (Operation Status Enable Register) (see page 86

Pagina 79

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 17 :SBUS<n>:UART:COUNt:RXFRames 543:SBUS<n>:UART:COUNt:TXFRames

Pagina 80

170 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands The :OPEE? query returns the current value contained

Pagina 81

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 171 :OPERegister:CONDition (Operation Status Condition Re

Pagina 82

172 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands • ":OPERegister[:EVENt] (Operation Status Event

Pagina 83

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 173 :OPERegister[:EVENt] (Operation Status Event Register

Pagina 84

174 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands • ":OPEE (Operation Status Enable Register)"

Pagina 85

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 175 :OVLenable (Overload Event Enable Register) (see page

Pagina 86

176 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands Query Syntax :OVLenable?The :OVLenable query returns

Pagina 87

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 177 :OVLRegister (Overload Event Register) (see page 860)

Pagina 88

178 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands <value> ::= integer in NR1 format.See Also • &q

Pagina 89

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 179 :PRINt (see page 860)Command Syntax:PRINt [<option

Pagina 90

18 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide :SEARch:SERial:CAN Commands 582:SEARch:SERial:CAN:PATTern:DATA 583:SEARch

Pagina 91

180 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands :RUN (see page 860)Command Syntax:RUNThe :RUN command

Pagina 92

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 181 :SERial (see page 860)Query Syntax:SERial?The :SERial

Pagina 93

182 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands :SINGle (see page 860)Command Syntax:SINGleThe :SINGl

Pagina 94

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 183 :STATus (see page 860)Query Syntax:STATus? <source

Pagina 95

184 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands :STOP (see page 860)Command Syntax:STOPThe :STOP comm

Pagina 96

Root (:) Commands 6Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 185 :TER (Trigger Event Register) (see page 860)Query Syn

Pagina 97

186 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide6 Root (:) Commands :VIEW (see page 860)Command Syntax:VIEW <source>

Pagina 98

187 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide7:ACQuire CommandsSet the parameters for acquiring and storing data. Se

Pagina 99

188 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide7 :ACQuire Commands The :ACQuire:TYPE NORMal command sets the oscilloscop

Pagina 100 - 4 Commands Quick Reference

:ACQuire Commands 7Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 189 :ACQuire:COMPlete (see page 860)Command Syntax:ACQuir

Pagina 101

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 19 :SYSTem:PROTection:LOCK 624:SYSTem:SETup 625:SYSTem:TIME 62727 :TIMebase

Pagina 102

190 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide7 :ACQuire Commands :ACQuire:COUNt (see page 860)Command Syntax:ACQuire:C

Pagina 103

:ACQuire Commands 7Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 191 :ACQuire:MODE (see page 860)Command Syntax:ACQuire:MO

Pagina 104

192 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide7 :ACQuire Commands :ACQuire:POINts (see page 860)Query Syntax:ACQuire:PO

Pagina 105

:ACQuire Commands 7Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 193 :ACQuire:SEGMented:ANALyze (see page 860)Command Synt

Pagina 106

194 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide7 :ACQuire Commands :ACQuire:SEGMented:COUNt (see page 860)Command Syntax

Pagina 107

:ACQuire Commands 7Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 195 :ACQuire:SEGMented:INDex (see page 860)Command Syntax

Pagina 108

196 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide7 :ACQuire Commands Option ExplicitPublic myMgr As VisaComLib.ResourceMan

Pagina 109

:ACQuire Commands 7Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 197 For lngI = lngSegments To 1 Step -1' Set the seg

Pagina 110

198 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide7 :ACQuire Commands :ACQuire:SRATe (see page 860)Query Syntax:ACQuire:SRA

Pagina 111

:ACQuire Commands 7Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 199 :ACQuire:TYPE (see page 860)Command Syntax:ACQuire:TY

Pagina 112

Notices© Agilent Technologies, Inc. 2005-2011No part of this manual may be reproduced in any form or by any means (including electronic storage and re

Pagina 113

20 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide :TRIGger:GLITch:QUALifier 668:TRIGger:GLITch:RANGe 669:TRIGger:GLITch:SOU

Pagina 114

200 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide7 :ACQuire Commands <acq_type> ::= {NORM | AVER | HRES | PEAK}See A

Pagina 115

201 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide8:BUS<n> CommandsControl all oscilloscope functions associated wi

Pagina 116

202 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide8 :BUS<n> Commands Introduction to:BUS<n>Commands<n> ::

Pagina 117

:BUS<n> Commands 8Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 203 :BUS<n>:BIT<m> (see page 860)Command

Pagina 118

204 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide8 :BUS<n> Commands :BUS<n>:BITS (see page 860)Command Syntax:

Pagina 119 - NR3 format

:BUS<n> Commands 8Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 205 ' Include digital channels 1 through 5, 8,

Pagina 120

206 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide8 :BUS<n> Commands :BUS<n>:CLEar (see page 860)Command Syntax

Pagina 121 - Syntax Elements

:BUS<n> Commands 8Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 207 :BUS<n>:DISPlay (see page 860)Command Synt

Pagina 122

208 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide8 :BUS<n> Commands :BUS<n>:LABel (see page 860)Command Syntax

Pagina 123 - Commands Quick Reference 4

:BUS<n> Commands 8Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 209 :BUS<n>:MASK (see page 860)Command Syntax:

Pagina 124

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 21 29 :WAVeform Commands:WAVeform:BYTeorder 715:WAVeform:COUNt 716:WAVeform:

Pagina 125 - Common (*) Commands

210 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide8 :BUS<n> Commands

Pagina 126 - 5 Common (*) Commands

211 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide9:CALibrate CommandsUtility commands for viewing calibration status and

Pagina 127

212 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide9 :CALibrate Commands • Determining the state of the calibration factor p

Pagina 128

:CALibrate Commands 9Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 213 :CALibrate:DATE (see page 860)Query Syntax:CALibrat

Pagina 129 - *CLS (Clear Status)

214 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide9 :CALibrate Commands :CALibrate:LABel (see page 860)Command Syntax:CALib

Pagina 130

:CALibrate Commands 9Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 215 :CALibrate:OUTPut (see page 860)Command Syntax:CALi

Pagina 131 - Common (*) Commands 5

216 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide9 :CALibrate Commands :CALibrate:PROTected (see page 860)Query Syntax:CAL

Pagina 132

:CALibrate Commands 9Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 217 :CALibrate:STARt (see page 860)Command Syntax:CALib

Pagina 133

218 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide9 :CALibrate Commands :CALibrate:STATus (see page 860)Query Syntax:CALibr

Pagina 134 - *IDN (Identification Number)

:CALibrate Commands 9Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 219 :CALibrate:TEMPerature (see page 860)Query Syntax:C

Pagina 135 - *LRN (Learn Device Setup)

22 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide :WMEMory<r>:SAVE 764:WMEMory<r>:SKEW 765:WMEMory<r>:YOF

Pagina 136 - *OPC (Operation Complete)

220 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide9 :CALibrate Commands :CALibrate:TIME (see page 860)Query Syntax:CALibrat

Pagina 137 - *OPT (Option Identification)

221 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10:CHANnel<n> CommandsControl all oscilloscope functions associat

Pagina 138

222 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10 :CHANnel<n> Commands Introduction to:CHANnel<n>Commands<

Pagina 139 - *RCL (Recall)

:CHANnel<n> Commands 10Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 223 The CHANnel<n> subsystem commands con

Pagina 140 - *RST (Reset)

224 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10 :CHANnel<n> Commands :CHANnel<n>:BWLimit (see page 860)Com

Pagina 141

:CHANnel<n> Commands 10Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 225 :CHANnel<n>:COUPling (see page 860)Co

Pagina 142

226 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10 :CHANnel<n> Commands :CHANnel<n>:DISPlay (see page 860)Com

Pagina 143 - *SAV (Save)

:CHANnel<n> Commands 10Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 227 :CHANnel<n>:IMPedance (see page 860)C

Pagina 144 - *SRE (Service Request Enable)

228 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10 :CHANnel<n> Commands :CHANnel<n>:INVert (see page 860)Comm

Pagina 145 - <mask><NL>

:CHANnel<n> Commands 10Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 229 :CHANnel<n>:LABel (see page 860)Comma

Pagina 146 - *STB (Read Status Byte)

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 23 :MTESt:RUMode 809:MTESt:RUMode:SOFailure 810:MTESt:{STARt | STOP} 811:MTE

Pagina 147

230 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10 :CHANnel<n> Commands :CHANnel<n>:OFFSet (see page 860)Comm

Pagina 148 - *TRG (Trigger)

:CHANnel<n> Commands 10Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 231 :CHANnel<n>:PROBe (see page 860)Comma

Pagina 149 - *TST (Self Test)

232 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10 :CHANnel<n> Commands :CHANnel<n>:PROBe:HEAD[:TYPE] (see pa

Pagina 150 - *WAI (Wait To Continue)

:CHANnel<n> Commands 10Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 233 :CHANnel<n>:PROBe:ID (see page 860)Qu

Pagina 151 - Root (:) Commands

234 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10 :CHANnel<n> Commands :CHANnel<n>:PROBe:SKEW (see page 860)

Pagina 152 - 6 Root (:) Commands

:CHANnel<n> Commands 10Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 235 :CHANnel<n>:PROBe:STYPe (see page 860

Pagina 153

236 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10 :CHANnel<n> Commands :CHANnel<n>:PROTection (see page 860)

Pagina 154

:CHANnel<n> Commands 10Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 237 :CHANnel<n>:RANGe (see page 860)Comma

Pagina 155 - :ACTivity

238 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10 :CHANnel<n> Commands :CHANnel<n>:SCALe (see page 860)Comma

Pagina 156 - :AER (Arm Event Register)

:CHANnel<n> Commands 10Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 239 :CHANnel<n>:UNITs (see page 860)Comma

Pagina 157 - :AUToscale

24 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide Blocking Synchronization 851Polling Synchronization With Timeout 852Synch

Pagina 158

240 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide10 :CHANnel<n> Commands :CHANnel<n>:VERNier (see page 860)Com

Pagina 159 - :AUToscale:AMODE

241 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide11:DEMO CommandsWhen the education kit is licensed (Option EDU), you ca

Pagina 160 - :AUToscale:CHANnels

242 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide11 :DEMO Commands :DEMO:FUNCtion (see page 860)Command Syntax:DEMO:FUNCti

Pagina 161 - :AUToscale:FDEBug

:DEMO Commands 11Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 243 MSO 3.1 kHz stair-step sine wave output of DAC @ ~1.5 V

Pagina 162

244 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide11 :DEMO Commands Query Syntax :DEMO:FUNCtion?The :DEMO:FUNCtion? query r

Pagina 163 - :DIGitize

:DEMO Commands 11Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 245 :DEMO:FUNCtion:PHASe:PHASe (see page 860)Command Syntax

Pagina 164

246 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide11 :DEMO Commands :DEMO:OUTPut (see page 860)Command Syntax:DEMO:OUTPut &

Pagina 165 - <value><NL>

247 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide12:DIGital<d> CommandsControl all oscilloscope functions associat

Pagina 166

248 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide12 :DIGital<d> Commands The DIGital subsystem commands control the

Pagina 167 - :MTERegister[:EVENt]?

:DIGital<d> Commands 12Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 249 :DIGital<d>:DISPlay (see page 860)Com

Pagina 168

25 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide1What's NewWhat's New in Version 1.10 26Version 1.00 at Introd

Pagina 169 - :OPEE <mask>

250 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide12 :DIGital<d> Commands :DIGital<d>:LABel (see page 860)Comma

Pagina 170

:DIGital<d> Commands 12Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 251 :DIGital<d>:POSition (see page 860)Co

Pagina 171 - :OPERegister:CONDition?

252 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide12 :DIGital<d> Commands :DIGital<d>:SIZE (see page 860)Comman

Pagina 172

:DIGital<d> Commands 12Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 253 :DIGital<d>:THReshold (see page 860)C

Pagina 173 - :OPERegister[:EVENt]?

254 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide12 :DIGital<d> Commands

Pagina 174

255 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide13:DISPlay CommandsControl how waveforms, graticule, and text are displ

Pagina 175

256 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide13 :DISPlay Commands • Save and Recall display data.Reporting the SetupUs

Pagina 176

:DISPlay Commands 13Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 257 :DISPlay:CLEar (see page 860)Command Syntax:DISPlay:

Pagina 177 - :OVLRegister?

258 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide13 :DISPlay Commands :DISPlay:DATA (see page 860)Query Syntax:DISPlay:DAT

Pagina 178

:DISPlay Commands 13Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 259 :DISPlay:LABel (see page 860)Command Syntax:DISPlay:

Pagina 179 - Root (:) Commands 6

26 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide1 What's New What's New in Version 1.10New command descriptions

Pagina 180

260 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide13 :DISPlay Commands :DISPlay:LABList (see page 860)Command Syntax:DISPla

Pagina 181 - Return Format:

:DISPlay Commands 13Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 261 :DISPlay:PERSistence (see page 860)Command Syntax:DI

Pagina 182

262 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide13 :DISPlay Commands :DISPlay:VECTors (see page 860)Command Syntax:DISPla

Pagina 183

263 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide14:EXTernal Trigger CommandsControl the input characteristics of the ex

Pagina 184

264 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide14 :EXTernal Trigger Commands :EXTernal:BWLimit (see page 860)Command Syn

Pagina 185 - :TER (Trigger Event Register)

:EXTernal Trigger Commands 14Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 265 :EXTernal:PROBe (see page 860)Command Synta

Pagina 186

266 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide14 :EXTernal Trigger Commands :EXTernal:RANGe (see page 860)Command Synta

Pagina 187 - :ACQuire Commands

:EXTernal Trigger Commands 14Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 267 :EXTernal:UNITs (see page 860)Command Synta

Pagina 188 - 7 :ACQuire Commands

268 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide14 :EXTernal Trigger Commands

Pagina 189 - :ACQuire:COMPlete

269 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide15:FUNCtion CommandsControl functions in the measurement/storage module

Pagina 190 - :ACQuire:COUNt

What's New 1Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 27 Version 1.00 at IntroductionThe Agilent InfiniiVision 30

Pagina 191 - :ACQuire:MODE

270 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide15 :FUNCtion Commands :FUNCtion:RANGe <range> (see page 279):FUNCti

Pagina 192 - :ACQuire:POINts

:FUNCtion Commands 15Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 271 Introduction to:FUNCtionCommandsThe FUNCtion subsys

Pagina 193 - :ACQuire:SEGMented:ANALyze

272 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide15 :FUNCtion Commands :FUNCtion:CENTer (see page 860)Command Syntax:FUNCt

Pagina 194 - :ACQuire:SEGMented:COUNt

:FUNCtion Commands 15Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 273 :FUNCtion:DISPlay (see page 860)Command Syntax:FUNC

Pagina 195 - :ACQuire:SEGMented:INDex

274 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide15 :FUNCtion Commands :FUNCtion:GOFT:OPERation (see page 860)Command Synt

Pagina 196

:FUNCtion Commands 15Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 275 :FUNCtion:GOFT:SOURce1 (see page 860)Command Syntax

Pagina 197 - :ACQuire Commands 7

276 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide15 :FUNCtion Commands :FUNCtion:GOFT:SOURce2 (see page 860)Command Syntax

Pagina 198 - :ACQuire:SRATe

:FUNCtion Commands 15Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 277 :FUNCtion:OFFSet (see page 860)Command Syntax:FUNCt

Pagina 199 - :ACQuire:TYPE

278 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide15 :FUNCtion Commands :FUNCtion:OPERation (see page 860)Command Syntax:FU

Pagina 200

:FUNCtion Commands 15Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 279 :FUNCtion:RANGe (see page 860)Command Syntax:FUNCti

Pagina 201 - :BUS<n> Commands

28 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide1 What's New Command Differences From 7000B Series OscilloscopesThe A

Pagina 202 - 8 :BUS<n> Commands

280 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide15 :FUNCtion Commands :FUNCtion:REFerence (see page 860)Command Syntax:FU

Pagina 203 - :BUS<n>:BIT<m>

:FUNCtion Commands 15Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 281 :FUNCtion:SCALe (see page 860)Command Syntax:FUNCti

Pagina 204 - :BUS<n>:BITS

282 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide15 :FUNCtion Commands :FUNCtion:SOURce1 (see page 860)Command Syntax:FUNC

Pagina 205 - :BUS<n> Commands 8

:FUNCtion Commands 15Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 283 :FUNCtion:SOURce2 (see page 860)Command Syntax:FUNC

Pagina 206 - :BUS<n>:CLEar

284 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide15 :FUNCtion Commands :FUNCtion:SPAN (see page 860)Command Syntax:FUNCtio

Pagina 207 - :BUS<n>:DISPlay

:FUNCtion Commands 15Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 285 :FUNCtion:WINDow (see page 860)Command Syntax:FUNCt

Pagina 208 - :BUS<n>:LABel

286 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide15 :FUNCtion Commands

Pagina 209 - :BUS<n>:MASK

287 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide16:HARDcopy CommandsSet and query the selection of hardcopy device and

Pagina 210

288 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide16 :HARDcopy Commands Introduction to:HARDcopyCommandsThe HARDcopy subsys

Pagina 211 - :CALibrate Commands

:HARDcopy Commands 16Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 289 :HARDcopy:AREA (see page 860)Command Syntax:HARDcop

Pagina 212 - 9 :CALibrate Commands

What's New 1Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 29 :MEASure:PPULses (see page 353)Counts the number of posi

Pagina 213 - :CALibrate:DATE

290 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide16 :HARDcopy Commands :HARDcopy:APRinter (see page 860)Command Syntax:HAR

Pagina 214 - :CALibrate:LABel

:HARDcopy Commands 16Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 291 :HARDcopy:FACTors (see page 860)Command Syntax:HARD

Pagina 215 - :CALibrate:OUTPut

292 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide16 :HARDcopy Commands :HARDcopy:FFEed (see page 860)Command Syntax:HARDco

Pagina 216 - :CALibrate:PROTected

:HARDcopy Commands 16Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 293 :HARDcopy:INKSaver (see page 860)Command Syntax:HAR

Pagina 217 - :CALibrate:STARt

294 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide16 :HARDcopy Commands :HARDcopy:LAYout (see page 860)Command Syntax:HARDc

Pagina 218 - :CALibrate:STATus

:HARDcopy Commands 16Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 295 :HARDcopy:NETWork:ADDRess (see page 860)Command Syn

Pagina 219 - :CALibrate:TEMPerature

296 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide16 :HARDcopy Commands :HARDcopy:NETWork:APPLy (see page 860)Command Synta

Pagina 220 - :CALibrate:TIME

:HARDcopy Commands 16Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 297 :HARDcopy:NETWork:DOMain (see page 860)Command Synt

Pagina 221 - :CHANnel<n> Commands

298 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide16 :HARDcopy Commands :HARDcopy:NETWork:PASSword (see page 860)Command Sy

Pagina 222

:HARDcopy Commands 16Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 299 :HARDcopy:NETWork:SLOT (see page 860)Command Syntax

Pagina 223 - :CHANnel<n> Commands 10

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 3In This BookThis book is your guide to programming the 3000 X- Series osci

Pagina 224 - :CHANnel<n>:BWLimit

30 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide1 What's New ChangedCommands:TRIGger:RUNT Commands (see page 680)Comm

Pagina 225 - :CHANnel<n>:COUPling

300 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide16 :HARDcopy Commands :HARDcopy:NETWork:USERname (see page 860)Command Sy

Pagina 226 - :CHANnel<n>:DISPlay

:HARDcopy Commands 16Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 301 :HARDcopy:PALette (see page 860)Command Syntax:HARD

Pagina 227 - :CHANnel<n>:IMPedance

302 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide16 :HARDcopy Commands :HARDcopy:PRINter:LIST (see page 860)Query Syntax:H

Pagina 228 - :CHANnel<n>:INVert

:HARDcopy Commands 16Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 303 :HARDcopy:STARt (see page 860)Command Syntax:HARDco

Pagina 229 - :CHANnel<n>:LABel

304 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide16 :HARDcopy Commands

Pagina 230 - :CHANnel<n>:OFFSet

305 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide17:LISTer CommandsIntroduction to:LISTerCommandsThe LISTer subsystem is

Pagina 231 - :CHANnel<n>:PROBe

306 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide17 :LISTer Commands :LISTer:DATA (see page 860)Query Syntax:LISTer:DATA?T

Pagina 232

:LISTer Commands 17Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 307 :LISTer:DISPlay (see page 860)Command Syntax:LISTer:D

Pagina 233 - :CHANnel<n>:PROBe:ID

308 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide17 :LISTer Commands :LISTer:REFerence (see page 860)Command Syntax:LISTer

Pagina 234 - :CHANnel<n>:PROBe:SKEW

309 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide18:MARKer CommandsSet and query the settings of X-axis markers (X1 and

Pagina 235 - :CHANnel<n>:PROBe:STYPe

What's New 1Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 31 ObsoleteCommandsDiscontinuedCommands:WAVeform:SOURce (se

Pagina 236 - :CHANnel<n>:PROTection

310 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide18 :MARKer Commands Introduction to:MARKerCommandsThe MARKer subsystem co

Pagina 237 - :CHANnel<n>:RANGe

:MARKer Commands 18Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 311 :MARKer:MODE (see page 860)Command Syntax:MARKer:MODE

Pagina 238 - :CHANnel<n>:SCALe

312 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide18 :MARKer Commands :MARKer:X1Position (see page 860)Command Syntax:MARKe

Pagina 239 - :CHANnel<n>:UNITs

:MARKer Commands 18Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 313 :MARKer:X1Y1source (see page 860)Command Syntax:MARKe

Pagina 240 - :CHANnel<n>:VERNier

314 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide18 :MARKer Commands :MARKer:X2Position (see page 860)Command Syntax:MARKe

Pagina 241 - :DEMO Commands

:MARKer Commands 18Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 315 :MARKer:X2Y2source (see page 860)Command Syntax:MARKe

Pagina 242 - :DEMO:FUNCtion

316 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide18 :MARKer Commands :MARKer:XDELta (see page 860)Query Syntax:MARKer:XDEL

Pagina 243

:MARKer Commands 18Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 317 :MARKer:Y1Position (see page 860)Command Syntax:MARKe

Pagina 244 - Demo Signal

318 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide18 :MARKer Commands :MARKer:Y2Position (see page 860)Command Syntax:MARKe

Pagina 245 - :DEMO:FUNCtion:PHASe:PHASe

:MARKer Commands 18Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 319 :MARKer:YDELta (see page 860)Query Syntax:MARKer:YDEL

Pagina 246 - :DEMO:OUTPut

32 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide1 What's New

Pagina 247 - :DIGital<d> Commands

320 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide18 :MARKer Commands

Pagina 248

321 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19:MEASure CommandsSelect automatic measurements to be made and control

Pagina 249 - :DIGital<d>:DISPlay

322 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:DEFine DELay, <delay spec> (see page

Pagina 250 - :DIGital<d>:LABel

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 323 :MEASure:FALLtime [<source>] (see page 343):ME

Pagina 251 - :DIGital<d>:POSition

324 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:NWIDth [<source>] (see page 347):MEAS

Pagina 252 - :DIGital<d>:SIZE

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 325 :MEASure:PHASe [<source1>] [,<source2>]

Pagina 253 - :DIGital<d>:THReshold

326 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:RISetime [<source>] (see page 359):ME

Pagina 254

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 327 n/a :MEASure:TEDGe? <slope><occurrence>[

Pagina 255 - :DISPlay Commands

328 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:VAMPlitude [<source>] (see page 371):

Pagina 256

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 329 :MEASure:VPP [<source>] (see page 376):MEASure

Pagina 257 - :DISPlay:CLEar

33 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide2Setting UpStep 1. Install Agilent IO Libraries Suite software 34Step 2.

Pagina 258 - :DISPlay:DATA

330 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands Introduction to:MEASureCommandsThe commands in the M

Pagina 259 - :DISPlay:LABel

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 331 If a measurement cannot be made (typically because t

Pagina 260 - :DISPlay:LABList

332 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:ALL (see page 860)Command Syntax:MEASure:AL

Pagina 261 - :DISPlay:PERSistence

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 333 :MEASure:AREa (see page 860)Command Syntax:MEASure:A

Pagina 262 - :DISPlay:VECTors

334 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:BWIDth (see page 860)Command Syntax:MEASure

Pagina 263 - :EXTernal Trigger Commands

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 335 :MEASure:CLEar (see page 860)Command Syntax:MEASure:

Pagina 264 - :EXTernal:BWLimit

336 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:COUNter (see page 860)Command Syntax:MEASur

Pagina 265 - :EXTernal:PROBe

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 337 :MEASure:DEFine (see page 860)Command Syntax:MEASure

Pagina 266 - :EXTernal:RANGe

338 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands This command defines the behavior of the :MEASure:DE

Pagina 267 - :EXTernal:UNITs

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 339 Return Format for <meas_spec> = DELay:{ <ed

Pagina 268

34 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide2 Setting Up Step 1. Install Agilent IO Libraries Suite software1 Download

Pagina 269 - :FUNCtion Commands

340 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:DELay (see page 860)Command Syntax:MEASure:

Pagina 270

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 341 Vtop. If you want to move the delay measurement poin

Pagina 271 - :FUNCtion Commands 15

342 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:DUTYcycle (see page 860)Command Syntax:MEAS

Pagina 272 - :FUNCtion:CENTer

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 343 :MEASure:FALLtime (see page 860)Command Syntax:MEASu

Pagina 273 - :FUNCtion:DISPlay

344 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:FREQuency (see page 860)Command Syntax:MEAS

Pagina 274 - :FUNCtion:GOFT:OPERation

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 345 :MEASure:NEDGes (see page 860)Command Syntax:MEASure

Pagina 275 - :FUNCtion:GOFT:SOURce1

346 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:NPULses (see page 860)Command Syntax:MEASur

Pagina 276 - :FUNCtion:GOFT:SOURce2

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 347 :MEASure:NWIDth (see page 860)Command Syntax:MEASure

Pagina 277 - :FUNCtion:OFFSet

348 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:OVERshoot (see page 860)Command Syntax:MEAS

Pagina 278 - :FUNCtion:OPERation

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 349 • ":MEASure:VTOP" on page 380• ":MEA

Pagina 279 - :FUNCtion:RANGe

Setting Up 2Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 35 Step 2. Connect and set up the oscilloscopeThe 3000 X-Series

Pagina 280 - :FUNCtion:REFerence

350 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:PEDGes (see page 860)Command Syntax:MEASure

Pagina 281 - :FUNCtion:SCALe

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 351 :MEASure:PERiod (see page 860)Command Syntax:MEASure

Pagina 282 - :FUNCtion:SOURce1

352 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:PHASe (see page 860)Command Syntax:MEASure:

Pagina 283 - :FUNCtion:SOURce2

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 353 :MEASure:PPULses (see page 860)Command Syntax:MEASur

Pagina 284 - :FUNCtion:SPAN

354 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:PREShoot (see page 860)Command Syntax:MEASu

Pagina 285 - :FUNCtion:WINDow

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 355 :MEASure:PWIDth (see page 860)Command Syntax:MEASure

Pagina 286

356 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:RESults (see page 860)Query Syntax:MEASure:

Pagina 287 - :HARDcopy Commands

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 357 Private Declare Sub Sleep Lib "kernel32" (

Pagina 288

358 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands ValueColumnArray(5) = "Std_Dev"ValueColumn

Pagina 289 - :HARDcopy:AREA

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 359 :MEASure:RISetime (see page 860)Command Syntax:MEASu

Pagina 290 - :HARDcopy:APRinter

36 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide2 Setting Up If automatic configuration is not supported, get the oscillos

Pagina 291 - :HARDcopy:FACTors

360 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:SDEViation (see page 860)Command Syntax:MEA

Pagina 292 - :HARDcopy:FFEed

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 361 :MEASure:SHOW (see page 860)Command Syntax:MEASure:S

Pagina 293 - :HARDcopy:INKSaver

362 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:SOURce (see page 860)Command Syntax:MEASure

Pagina 294 - :HARDcopy:LAYout

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 363 • ":MEASure:PHASe" on page 352Example Cod

Pagina 295 - :HARDcopy:NETWork:ADDRess

364 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:STATistics (see page 860)Command Syntax:MEA

Pagina 296 - :HARDcopy:NETWork:APPLy

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 365 :MEASure:STATistics:INCRement (see page 860)Command

Pagina 297 - :HARDcopy:NETWork:DOMain

366 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:STATistics:RESet (see page 860)Command Synt

Pagina 298 - :HARDcopy:NETWork:PASSword

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 367 :MEASure:TEDGe (see page 860)Query Syntax:MEASure:TE

Pagina 299 - :HARDcopy:NETWork:SLOT

368 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands Return Format <value><NL><value> :

Pagina 300 - :HARDcopy:NETWork:USERname

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 369 :MEASure:TVALue (see page 860)Query Syntax:MEASure:T

Pagina 301 - :HARDcopy:PALette

Setting Up 2Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 37 Step 3. Verify the oscilloscope connection1 On the controller

Pagina 302 - :HARDcopy:PRINter:LIST

370 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands <value> ::= time in seconds of the specified v

Pagina 303 - :HARDcopy:STARt

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 371 :MEASure:VAMPlitude (see page 860)Command Syntax:MEA

Pagina 304

372 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:VAVerage (see page 860)Command Syntax:MEASu

Pagina 305 - :LISTer Commands

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 373 :MEASure:VBASe (see page 860)Command Syntax:MEASure:

Pagina 306 - :LISTer:DATA

374 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:VMAX (see page 860)Command Syntax:MEASure:V

Pagina 307 - :LISTer:DISPlay

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 375 :MEASure:VMIN (see page 860)Command Syntax:MEASure:V

Pagina 308 - :LISTer:REFerence

376 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:VPP (see page 860)Command Syntax:MEASure:VP

Pagina 309 - :MARKer Commands

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 377 :MEASure:VRATio (see page 860)Command Syntax:MEASure

Pagina 310

378 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:VRMS (see page 860)Command Syntax:MEASure:V

Pagina 311 - :MARKer:MODE

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 379 :MEASure:VTIMe (see page 860)Query Syntax:MEASure:VT

Pagina 312 - :MARKer:X1Position

38 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide2 Setting Up You must manually add instruments on LAN interfaces:a Right-c

Pagina 313 - :MARKer:X1Y1source

380 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:VTOP (see page 860)Command Syntax:MEASure:V

Pagina 314 - :MARKer:X2Position

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 381 :MEASure:WINDow (see page 860)Command Syntax:MEASure

Pagina 315 - :MARKer:X2Y2source

382 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands :MEASure:XMAX (see page 860)Command Syntax:MEASure:X

Pagina 316 - :MARKer:XDELta

:MEASure Commands 19Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 383 :MEASure:XMIN (see page 860)Command Syntax:MEASure:X

Pagina 317 - :MARKer:Y1Position

384 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide19 :MEASure Commands

Pagina 318 - :MARKer:Y2Position

385 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20:MTESt CommandsThe MTESt subsystem commands and queries control the m

Pagina 319 - :MARKer:YDELta

386 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:DELete (see page 401)n/a n/a:MTESt:ENABle {{0 |

Pagina 320

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 387 Introduction to:MTEStCommandsMask testing automaticall

Pagina 321 - :MEASure Commands

388 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands Private Declare Sub Sleep Lib "kernel32" (By

Pagina 322

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 389 myScope.WriteString ":MTESt:AMASk:CREate"Deb

Pagina 323

Setting Up 2Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 39 Otherwise, if the instrument is not on the same subnet, click

Pagina 324

390 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:ALL (see page 860)Command Syntax:MTESt:ALL <

Pagina 325

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 391 :MTESt:AMASk:CREate (see page 860)Command Syntax:MTESt

Pagina 326

392 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:AMASk:SOURce (see page 860)Command Syntax:MTESt

Pagina 327

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 393 :MTESt:AMASk:UNITs (see page 860)Command Syntax:MTESt:

Pagina 328

394 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:AMASk:XDELta (see page 860)Command Syntax:MTESt

Pagina 329

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 395 :MTESt:AMASk:YDELta (see page 860)Command Syntax:MTESt

Pagina 330

396 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:COUNt:FWAVeforms (see page 860)Query Syntax:MTE

Pagina 331 - :MEASure Commands 19

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 397 :MTESt:COUNt:RESet (see page 860)Command Syntax:MTESt:

Pagina 332 - :MEASure:ALL

398 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:COUNt:TIME (see page 860)Query Syntax:MTESt:COU

Pagina 333 - :MEASure:AREa

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 399 :MTESt:COUNt:WAVeforms (see page 860)Query Syntax:MTES

Pagina 334 - :MEASure:BWIDth

4 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide• Chapter 9, “:CALibrate Commands,” starting on page 211, describes utilit

Pagina 335 - :MEASure:CLEar

40 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide2 Setting Up iii If the instrument is successfully opened, click OK to clo

Pagina 336 - :MEASure:COUNter

400 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:DATA (see page 860)Command Syntax:MTESt:DATA &l

Pagina 337 - :MEASure:DEFine

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 401 :MTESt:DELete (see page 860)Command Syntax:MTESt:DELet

Pagina 338

402 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:ENABle (see page 860)Command Syntax:MTESt:ENABl

Pagina 339

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 403 :MTESt:LOCK (see page 860)Command Syntax:MTESt:LOCK &l

Pagina 340 - :MEASure:DELay

404 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:RMODe (see page 860)Command Syntax:MTESt:RMODe

Pagina 341

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 405 :MTESt:RMODe:FACTion:MEASure (see page 860)Command Syn

Pagina 342 - :MEASure:DUTYcycle

406 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:RMODe:FACTion:PRINt (see page 860)Command Synta

Pagina 343 - :MEASure:FALLtime

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 407 :MTESt:RMODe:FACTion:SAVE (see page 860)Command Syntax

Pagina 344 - :MEASure:FREQuency

408 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:RMODe:FACTion:STOP (see page 860)Command Syntax

Pagina 345 - :MEASure:NEDGes

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 409 :MTESt:RMODe:SIGMa (see page 860)Command Syntax:MTESt:

Pagina 346 - :MEASure:NPULses

Setting Up 2Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 41 3 Test some commands on the instrument:a Right- click on the

Pagina 347 - :MEASure:NWIDth

410 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:RMODe:TIME (see page 860)Command Syntax:MTESt:R

Pagina 348 - :MEASure:OVERshoot

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 411 :MTESt:RMODe:WAVeforms (see page 860)Command Syntax:MT

Pagina 349

412 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:SCALe:BIND (see page 860)Command Syntax:MTESt:S

Pagina 350 - :MEASure:PEDGes

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 413 :MTESt:SCALe:X1 (see page 860)Command Syntax:MTESt:SCA

Pagina 351 - :MEASure:PERiod

414 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:SCALe:XDELta (see page 860)Command Syntax:MTESt

Pagina 352 - :MEASure:PHASe

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 415 :MTESt:SCALe:Y1 (see page 860)Command Syntax:MTESt:SCA

Pagina 353 - :MEASure:PPULses

416 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:SCALe:Y2 (see page 860)Command Syntax:MTESt:SCA

Pagina 354 - :MEASure:PREShoot

:MTESt Commands 20Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 417 :MTESt:SOURce (see page 860)Command Syntax:MTESt:SOURc

Pagina 355 - :MEASure:PWIDth

418 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide20 :MTESt Commands :MTESt:TITLe (see page 860)Query Syntax:MTESt:TITLe?Th

Pagina 356 - :MEASure:RESults

419 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide21:POD CommandsControl all oscilloscope functions associated with group

Pagina 357

42 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide2 Setting Up

Pagina 358

420 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide21 :POD Commands :POD1:DISP 0;THR +1.40E+00

Pagina 359 - :MEASure:RISetime

:POD Commands 21Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 421 :POD<n>:DISPlay (see page 860)Command Syntax:POD&l

Pagina 360 - :MEASure:SDEViation

422 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide21 :POD Commands :POD<n>:SIZE (see page 860)Command Syntax:POD<n

Pagina 361 - :MEASure:SHOW

:POD Commands 21Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 423 :POD<n>:THReshold (see page 860)Command Syntax:POD

Pagina 362 - :MEASure:SOURce

424 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide21 :POD Commands ' Set channels 0-7 to CMOS threshold.myScope.WriteS

Pagina 363 - Example Code

425 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide22:RECall CommandsRecall previously saved oscilloscope setups, referenc

Pagina 364 - :MEASure:STATistics

426 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide22 :RECall Commands Return FormatThe following is a sample response from

Pagina 365 - :MEASure:STATistics:INCRement

:RECall Commands 22Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 427 :RECall:FILename (see page 860)Command Syntax:RECall:

Pagina 366 - :MEASure:STATistics:RESet

428 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide22 :RECall Commands :RECall:MASK[:STARt] (see page 860)Command Syntax:REC

Pagina 367 - :MEASure:TEDGe

:RECall Commands 22Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 429 :RECall:PWD (see page 860)Command Syntax:RECall:PWD &

Pagina 368

43 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide3Getting StartedBasic Oscilloscope Program Structure 44Programming the O

Pagina 369 - :MEASure:TVALue

430 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide22 :RECall Commands :RECall:SETup[:STARt] (see page 860)Command Syntax:RE

Pagina 370

:RECall Commands 22Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 431 :RECall:WMEMory<r>[:STARt] (see page 860)Comman

Pagina 371 - :MEASure:VAMPlitude

432 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide22 :RECall Commands

Pagina 372 - :MEASure:VAVerage

433 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23:SAVE CommandsSave oscilloscope setups, screen images, and data. See

Pagina 373 - :MEASure:VBASe

434 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23 :SAVE Commands Introduction to:SAVE CommandsThe :SAVE subsystem provid

Pagina 374 - :MEASure:VMAX

:SAVE Commands 23Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 435 The following is a sample response from the :SAVE? quer

Pagina 375 - :MEASure:VMIN

436 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23 :SAVE Commands :SAVE:FILename (see page 860)Command Syntax:SAVE:FILena

Pagina 376 - :MEASure:VPP

:SAVE Commands 23Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 437 :SAVE:IMAGe[:STARt] (see page 860)Command Syntax:SAVE:I

Pagina 377 - :MEASure:VRATio

438 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23 :SAVE Commands :SAVE:IMAGe:FACTors (see page 860)Command Syntax:SAVE:I

Pagina 378 - :MEASure:VRMS

:SAVE Commands 23Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 439 :SAVE:IMAGe:FORMat (see page 860)Command Syntax:SAVE:IM

Pagina 379 - :MEASure:VTIMe

44 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide3 Getting Started Basic Oscilloscope Program StructureThe following figure

Pagina 380 - :MEASure:VTOP

440 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23 :SAVE Commands :SAVE:IMAGe:INKSaver (see page 860)Command Syntax:SAVE:

Pagina 381 - :MEASure:WINDow

:SAVE Commands 23Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 441 :SAVE:IMAGe:PALette (see page 860)Command Syntax:SAVE:I

Pagina 382 - :MEASure:XMAX

442 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23 :SAVE Commands :SAVE:LISTer[:STARt] (see page 860)Command Syntax:SAVE:

Pagina 383 - :MEASure:XMIN

:SAVE Commands 23Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 443 :SAVE:MASK[:STARt] (see page 860)Command Syntax:SAVE:MA

Pagina 384

444 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23 :SAVE Commands :SAVE:PWD (see page 860)Command Syntax:SAVE:PWD <pat

Pagina 385 - :MTESt Commands

:SAVE Commands 23Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 445 :SAVE:SETup[:STARt] (see page 860)Command Syntax:SAVE:S

Pagina 386

446 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23 :SAVE Commands :SAVE:WAVeform[:STARt] (see page 860)Command Syntax:SAV

Pagina 387

:SAVE Commands 23Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 447 :SAVE:WAVeform:FORMat (see page 860)Command Syntax:SAVE

Pagina 388

448 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23 :SAVE Commands :SAVE:WAVeform:LENGth (see page 860)Command Syntax:SAVE

Pagina 389 - :MTESt Commands 20

:SAVE Commands 23Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 449 :SAVE:WAVeform:SEGMented (see page 860)Command Syntax:S

Pagina 390 - :MTESt:ALL

Getting Started 3Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 45 acquisition memory in the oscilloscope, or transferred t

Pagina 391 - :MTESt:AMASk:CREate

450 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23 :SAVE Commands :SAVE:WMEMory:SOURce (see page 860)Command Syntax:SAVE:

Pagina 392 - :MTESt:AMASk:SOURce

:SAVE Commands 23Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 451 :SAVE:WMEMory[:STARt] (see page 860)Command Syntax:SAVE

Pagina 393 - :MTESt:AMASk:UNITs

452 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide23 :SAVE Commands

Pagina 394 - :MTESt:AMASk:XDELta

453 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24:SBUS<n> CommandsControl the modes and parameters for each seri

Pagina 395 - :MTESt:AMASk:YDELta

454 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands • LIN (Local Interconnect Network) triggering—

Pagina 396 - :MTESt:COUNt:FWAVeforms

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 455 General :SBUS<n> CommandsTable 84 Genera

Pagina 397 - :MTESt:COUNt:RESet

456 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:DISPlay (see page 860)Command S

Pagina 398 - :MTESt:COUNt:TIME

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 457 :SBUS<n>:MODE (see page 860)Command Synt

Pagina 399 - :MTESt:COUNt:WAVeforms

458 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:CAN CommandsNOTEThese commands

Pagina 400 - :MTESt:DATA

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 459 :SBUS<n>:CAN:TRIGger:PATTern:DATA <st

Pagina 401 - :MTESt:DELete

46 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide3 Getting Started Programming the Oscilloscope• "Referencing the IO L

Pagina 402 - :MTESt:ENABle

460 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:CAN:COUNt:ERRor (see page 860)Q

Pagina 403 - :MTESt:LOCK

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 461 :SBUS<n>:CAN:COUNt:OVERload (see page 86

Pagina 404 - :MTESt:RMODe

462 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:CAN:COUNt:RESet (see page 860)C

Pagina 405 - :MTESt:RMODe:FACTion:MEASure

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 463 :SBUS<n>:CAN:COUNt:TOTal (see page 860)Q

Pagina 406 - :MTESt:RMODe:FACTion:PRINt

464 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:CAN:COUNt:UTILization (see page

Pagina 407 - :MTESt:RMODe:FACTion:SAVE

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 465 :SBUS<n>:CAN:SAMPlepoint (see page 860)C

Pagina 408 - :MTESt:RMODe:FACTion:STOP

466 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:CAN:SIGNal:BAUDrate (see page 8

Pagina 409 - :MTESt:RMODe:SIGMa

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 467 :SBUS<n>:CAN:SIGNal:DEFinition (see page

Pagina 410 - :MTESt:RMODe:TIME

468 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:CAN:SOURce (see page 860)Comman

Pagina 411 - :MTESt:RMODe:WAVeforms

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 469 :SBUS<n>:CAN:TRIGger (see page 860)Comma

Pagina 412 - :MTESt:SCALe:BIND

Getting Started 3Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 47 3 Click OK.To reference the Agilent VISA COM library in

Pagina 413 - :MTESt:SCALe:X1

470 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands CAN Id specification is set by the :SBUS<n&

Pagina 414 - :MTESt:SCALe:XDELta

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 471 :SBUS<n>:CAN:TRIGger:PATTern:DATA (see p

Pagina 415 - :MTESt:SCALe:Y1

472 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth

Pagina 416 - :MTESt:SCALe:Y2

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 473 :SBUS<n>:CAN:TRIGger:PATTern:ID (see pag

Pagina 417 - :MTESt:SOURce

474 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:CAN:TRIGger:PATTern:ID:MODE (se

Pagina 418 - :MTESt:TITLe

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 475 :SBUS<n>:I2S CommandsNOTEThese commands

Pagina 419 - :POD Commands

476 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:I2S:TRIGger <operator> (s

Pagina 420

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 477 :SBUS<n>:I2S:ALIGnment (see page 860)Com

Pagina 421 - :POD<n>:DISPlay

478 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:I2S:BASE (see page 860)Command

Pagina 422 - :POD<n>:SIZE

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 479 :SBUS<n>:I2S:CLOCk:SLOPe (see page 860)C

Pagina 423 - :POD<n>:THReshold

48 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide3 Getting Started Dim myMgr As VisaComLib.ResourceManagerDim myScope As Vi

Pagina 424

480 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:I2S:RWIDth (see page 860)Comman

Pagina 425 - :RECall Commands

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 481 :SBUS<n>:I2S:SOURce:CLOCk (see page 860)

Pagina 426

482 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:I2S:SOURce:DATA (see page 860)C

Pagina 427 - :RECall:FILename

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 483 :SBUS<n>:I2S:SOURce:WSELect (see page 86

Pagina 428 - :RECall:MASK[:STARt]

484 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:I2S:TRIGger (see page 860)Comma

Pagina 429 - :RECall:PWD

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 485 See Also • "Introduction to :TRIGger Comm

Pagina 430 - :RECall:SETup[:STARt]

486 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:I2S:TRIGger:AUDio (see page 860

Pagina 431 - :RECall Commands 22

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 487 :SBUS<n>:I2S:TRIGger:PATTern:DATA (see p

Pagina 432

488 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands Query Syntax :SBUS<n>:I2S:TRIGger:PATTer

Pagina 433 - :SAVE Commands

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 489 :SBUS<n>:I2S:TRIGger:PATTern:FORMat (see

Pagina 434

Getting Started 3Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 49 Vertical is set to 16 V full- scale (2 V/div) with cente

Pagina 435 - :SAVE Commands 23

490 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:I2S:TRIGger:RANGe (see page 860

Pagina 436 - :SAVE:FILename

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 491 • ":SBUS<n>:I2S:TRIGger" on p

Pagina 437 - :SAVE:IMAGe[:STARt]

492 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:I2S:TWIDth (see page 860)Comman

Pagina 438 - :SAVE:IMAGe:FACTors

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 493 :SBUS<n>:I2S:WSLow (see page 860)Command

Pagina 439 - :SAVE:IMAGe:FORMat

494 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:IIC CommandsNOTEThese commands

Pagina 440 - :SAVE:IMAGe:INKSaver

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 495 :SBUS<n>:IIC:ASIZe (see page 860)Command

Pagina 441 - :SAVE:IMAGe:PALette

496 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:IIC[:SOURce]:CLOCk (see page 86

Pagina 442 - :SAVE:LISTer[:STARt]

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 497 :SBUS<n>:IIC[:SOURce]:DATA (see page 860

Pagina 443 - :SAVE:MASK[:STARt]

498 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:IIC:TRIGger:PATTern:ADDRess (se

Pagina 444 - :SAVE:PWD

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 499 :SBUS<n>:IIC:TRIGger:PATTern:DATA (see p

Pagina 445 - :SAVE:SETup[:STARt]

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 5• Chapter 24, “:SBUS<n> Commands,” starting on page 453, describes c

Pagina 446 - :SAVE:WAVeform[:STARt]

50 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide3 Getting Started When you send the :DIGitize command to the oscilloscope,

Pagina 447 - :SAVE:WAVeform:FORMat

500 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:IIC:TRIGger:PATTern:DATa2 (see

Pagina 448 - :SAVE:WAVeform:LENGth

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 501 :SBUS<n>:IIC:TRIGger:QUALifier (see page

Pagina 449 - :SAVE:WAVeform:SEGMented

502 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:IIC:TRIGger[:TYPE] (see page 86

Pagina 450 - :SAVE:WMEMory:SOURce

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 503 • ":SBUS<n>:IIC:TRIGger:PATTern:ADD

Pagina 451 - :SAVE:WMEMory[:STARt]

504 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:LIN CommandsNOTEThese commands

Pagina 452

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 505 :SBUS<n>:LIN:TRIGger:PATTern:DATA <st

Pagina 453 - :SBUS<n> Commands

506 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:LIN:PARity (see page 860)Comman

Pagina 454

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 507 :SBUS<n>:LIN:SAMPlepoint (see page 860)C

Pagina 455

508 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:LIN:SIGNal:BAUDrate (see page 8

Pagina 456 - :SBUS<n>:DISPlay

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 509 :SBUS<n>:LIN:SOURce (see page 860)Comman

Pagina 457 - :SBUS<n>:MODE

Getting Started 3Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 51 The easiest method of transferring a digitized waveform

Pagina 458 - :SBUS<n>:CAN Commands

510 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:LIN:STANdard (see page 860)Comm

Pagina 459

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 511 :SBUS<n>:LIN:SYNCbreak (see page 860)Com

Pagina 460

512 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:LIN:TRIGger (see page 860)Comma

Pagina 461 - :SBUS<n> Commands 24

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 513 :SBUS<n>:LIN:TRIGger:ID (see page 860)Co

Pagina 462

514 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:LIN:TRIGger:PATTern:DATA (see p

Pagina 463

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 515 See Also • "Introduction to :TRIGger Comm

Pagina 464

516 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth

Pagina 465

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 517 :SBUS<n>:LIN:TRIGger:PATTern:FORMat (see

Pagina 466

518 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:SPI CommandsNOTEThese commands

Pagina 467

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 519 :SBUS<n>:SPI:SOURce:MISO <source>

Pagina 468 - :SBUS<n>:CAN:SOURce

52 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide3 Getting Started Reading Query Results into String VariablesThe output of

Pagina 469 - :SBUS<n>:CAN:TRIGger

520 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:SPI:BITorder (see page 860)Comm

Pagina 470

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 521 :SBUS<n>:SPI:CLOCk:SLOPe (see page 860)C

Pagina 471

522 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:SPI:CLOCk:TIMeout (see page 860

Pagina 472

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 523 :SBUS<n>:SPI:FRAMing (see page 860)Comma

Pagina 473

524 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:SPI:SOURce:CLOCk (see page 860)

Pagina 474

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 525 :SBUS<n>:SPI:SOURce:DATA (see page 860)C

Pagina 475 - :SBUS<n>:I2S Commands

526 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:SPI:SOURce:FRAMe (see page 860)

Pagina 476

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 527 :SBUS<n>:SPI:SOURce:MISO (see page 860)C

Pagina 477 - :SBUS<n>:I2S:ALIGnment

528 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:SPI:SOURce:MOSI (see page 860)C

Pagina 478 - :SBUS<n>:I2S:BASE

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 529 :SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA (

Pagina 479

Getting Started 3Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 53 The "8" states the number of digits that follo

Pagina 480 - :SBUS<n>:I2S:RWIDth

530 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh

Pagina 481

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 531 :SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA (

Pagina 482

532 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh

Pagina 483

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 533 :SBUS<n>:SPI:TRIGger:TYPE (see page 860)

Pagina 484 - :SBUS<n>:I2S:TRIGger

534 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:SPI:WIDTh (see page 860)Command

Pagina 485

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 535 :SBUS<n>:UART CommandsNOTEThese commands

Pagina 486

536 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:SOURce:RX <source> (

Pagina 487

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 537 :SBUS<n>:UART:TRIGger:TYPE <value>

Pagina 488

538 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:BASE (see page 860)Command

Pagina 489

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 539 :SBUS<n>:UART:BAUDrate (see page 860)Com

Pagina 490

54 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide3 Getting Started To read the :TIMebase:RANGe?;DELay? query result into mu

Pagina 491

540 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:BITorder (see page 860)Com

Pagina 492 - :SBUS<n>:I2S:TWIDth

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 541 :SBUS<n>:UART:COUNt:ERRor (see page 860)

Pagina 493 - :SBUS<n>:I2S:WSLow

542 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:COUNt:RESet (see page 860)

Pagina 494 - :SBUS<n>:IIC Commands

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 543 :SBUS<n>:UART:COUNt:RXFRames (see page 8

Pagina 495 - :SBUS<n>:IIC:ASIZe

544 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:COUNt:TXFRames (see page 8

Pagina 496

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 545 :SBUS<n>:UART:FRAMing (see page 860)Comm

Pagina 497

546 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:PARity (see page 860)Comma

Pagina 498

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 547 :SBUS<n>:UART:POLarity (see page 860)Com

Pagina 499

548 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:SOURce:RX (see page 860)Co

Pagina 500

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 549 :SBUS<n>:UART:SOURce:TX (see page 860)Co

Pagina 501

Getting Started 3Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 55 Other Ways of Sending CommandsStandard Commands for Prog

Pagina 502

550 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:TRIGger:BASE (see page 860

Pagina 503

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 551 :SBUS<n>:UART:TRIGger:BURSt (see page 86

Pagina 504 - :SBUS<n>:LIN Commands

552 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:TRIGger:DATA (see page 860

Pagina 505

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 553 :SBUS<n>:UART:TRIGger:IDLE (see page 860

Pagina 506 - :SBUS<n>:LIN:PARity

554 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:TRIGger:QUALifier (see pag

Pagina 507

:SBUS<n> Commands 24Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 555 :SBUS<n>:UART:TRIGger:TYPE (see page 860

Pagina 508

556 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide24 :SBUS<n> Commands :SBUS<n>:UART:WIDTh (see page 860)Comman

Pagina 509 - :SBUS<n>:LIN:SOURce

557 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25:SEARch CommandsControl the event search modes and parameters for eac

Pagina 510 - :SBUS<n>:LIN:STANdard

558 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands General :SEARch CommandsTable 91 General :SEARch Comm

Pagina 511 - :SBUS<n>:LIN:SYNCbreak

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 559 :SEARch:COUNt (see page 860)Query Syntax:SEARch:COUNt

Pagina 512 - :SBUS<n>:LIN:TRIGger

56 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide3 Getting Started

Pagina 513 - :SBUS<n>:LIN:TRIGger:ID

560 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:MODE (see page 860)Command Syntax:SEARch:MODE

Pagina 514

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 561 :SEARch:STATe (see page 860)Command Syntax:SEARch:STA

Pagina 515

562 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:EDGE CommandsTable 92 :SEARch:EDGE Commands S

Pagina 516

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 563 :SEARch:EDGE:SLOPe (see page 860)Command Syntax:SEARc

Pagina 517

564 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:EDGE:SOURce (see page 860)Command Syntax:SEAR

Pagina 518 - :SBUS<n>:SPI Commands

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 565 :SEARch:GLITch CommandsTable 93 :SEARch:GLITch Comman

Pagina 519

566 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:GLITch:GREaterthan (see page 860)Command Synt

Pagina 520 - :SBUS<n>:SPI:BITorder

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 567 :SEARch:GLITch:LESSthan (see page 860)Command Syntax:

Pagina 521

568 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:GLITch:POLarity (see page 860)Command Syntax:

Pagina 522

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 569 :SEARch:GLITch:QUALifier (see page 860)Command Syntax

Pagina 523 - :SBUS<n>:SPI:FRAMing

57 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4Commands Quick ReferenceCommand Summary 58Syntax Elements 121

Pagina 524

570 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:GLITch:RANGe (see page 860)Command Syntax:SEA

Pagina 525

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 571 :SEARch:GLITch:SOURce (see page 860)Command Syntax:SE

Pagina 526

572 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:RUNT CommandsTable 94 :SEARch:RUNT Commands S

Pagina 527

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 573 :SEARch:RUNT:POLarity (see page 860)Command Syntax:SE

Pagina 528

574 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:RUNT:QUALifier (see page 860)Command Syntax:S

Pagina 529

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 575 :SEARch:RUNT:SOURce (see page 860)Command Syntax:SEAR

Pagina 530

576 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:RUNT:TIME (see page 860)Command Syntax:SEARch

Pagina 531

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 577 :SEARch:TRANsition CommandsTable 95 :SEARch:TRANsitio

Pagina 532

578 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:TRANsition:QUALifier (see page 860)Command Sy

Pagina 533

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 579 :SEARch:TRANsition:SLOPe (see page 860)Command Syntax

Pagina 534 - :SBUS<n>:SPI:WIDTh

58 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference Command Summary• Common (*) Commands Summary (s

Pagina 535 - :SBUS<n>:UART Commands

580 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:TRANsition:SOURce (see page 860)Command Synta

Pagina 536

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 581 :SEARch:TRANsition:TIME (see page 860)Command Syntax:

Pagina 537

582 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:CAN CommandsTable 96 :SEARch:SERial:CA

Pagina 538 - :SBUS<n>:UART:BASE

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 583 :SEARch:SERial:CAN:PATTern:DATA (see page 860)Command

Pagina 539 - :SBUS<n>:UART:BAUDrate

584 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:CAN:PATTern:DATA:LENGth (see page 860)

Pagina 540 - :SBUS<n>:UART:BITorder

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 585 :SEARch:SERial:CAN:PATTern:ID (see page 860)Command S

Pagina 541

586 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:CAN:PATTern:ID:MODE (see page 860)Comm

Pagina 542

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 587 :SEARch:SERial:CAN:MODE (see page 860)Command Syntax:

Pagina 543

588 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:I2S CommandsTable 97 :SEARch:SERial:I2

Pagina 544

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 589 :SEARch:SERial:I2S:AUDio (see page 860)Command Syntax

Pagina 545 - :SBUS<n>:UART:FRAMing

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 59 • :SEARch:SERial:IIC Commands Summary (see page

Pagina 546 - :SBUS<n>:UART:PARity

590 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:I2S:MODE (see page 860)Command Syntax:

Pagina 547 - :SBUS<n>:UART:POLarity

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 591 :SEARch:SERial:I2S:PATTern:DATA (see page 860)Command

Pagina 548 - :SBUS<n>:UART:SOURce:RX

592 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:I2S:PATTern:FORMat (see page 860)Comma

Pagina 549 - :SBUS<n>:UART:SOURce:TX

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 593 :SEARch:SERial:I2S:RANGe (see page 860)Command Syntax

Pagina 550

594 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:IIC CommandsTable 98 :SEARch:SERial:II

Pagina 551

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 595 :SEARch:SERial:IIC:MODE (see page 860)Command Syntax:

Pagina 552

596 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands • ":SEARch:SERial:IIC:PATTern:ADDRess" on

Pagina 553

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 597 :SEARch:SERial:IIC:PATTern:ADDRess (see page 860)Comm

Pagina 554

598 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:IIC:PATTern:DATA (see page 860)Command

Pagina 555

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 599 :SEARch:SERial:IIC:PATTern:DATA2 (see page 860)Comman

Pagina 556 - :SBUS<n>:UART:WIDTh

6 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide• Chapter 37, “Programming Examples,” starting on page 869.Mixed-SignalOsc

Pagina 557 - :SEARch Commands

60 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference n/a *IDN? (see page 132) AGILENT TECHNOLOGIES,&

Pagina 558 - General :SEARch Commands

600 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:IIC:QUALifier (see page 860)Command Sy

Pagina 559 - :SEARch:COUNt

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 601 :SEARch:SERial:LIN CommandsTable 99 :SEARch:SERial:LI

Pagina 560 - :SEARch:MODE

602 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:LIN:ID (see page 860)Command Syntax:SE

Pagina 561 - :SEARch:STATe

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 603 :SEARch:SERial:LIN:MODE (see page 860)Command Syntax:

Pagina 562 - :SEARch:EDGE Commands

604 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:LIN:PATTern:DATA (see page 860)Command

Pagina 563 - :SEARch:EDGE:SLOPe

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 605 :SEARch:SERial:LIN:PATTern:DATA:LENGth (see page 860)

Pagina 564 - :SEARch:EDGE:SOURce

606 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:LIN:PATTern:FORMat (see page 860)Comma

Pagina 565 - :SEARch:GLITch Commands

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 607 :SEARch:SERial:SPI CommandsTable 100 :SEARch:SERial:S

Pagina 566 - :SEARch:GLITch:GREaterthan

608 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:SPI:MODE (see page 860)Command Syntax:

Pagina 567 - :SEARch:GLITch:LESSthan

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 609 :SEARch:SERial:SPI:PATTern:DATA (see page 860)Command

Pagina 568 - :SEARch:GLITch:POLarity

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 61 n/a *OPT? (see page 137) <return_value> :

Pagina 569 - :SEARch:GLITch:QUALifier

610 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:SPI:PATTern:WIDTh (see page 860)Comman

Pagina 570 - :SEARch:GLITch:RANGe

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 611 :SEARch:SERial:UART CommandsTable 101 :SEARch:SERial:

Pagina 571 - :SEARch:GLITch:SOURce

612 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:UART:DATA (see page 860)Command Syntax

Pagina 572 - :SEARch:RUNT Commands

:SEARch Commands 25Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 613 :SEARch:SERial:UART:MODE (see page 860)Command Syntax

Pagina 573 - :SEARch:RUNT:POLarity

614 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide25 :SEARch Commands :SEARch:SERial:UART:QUALifier (see page 860)Command S

Pagina 574 - :SEARch:RUNT:QUALifier

615 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide26:SYSTem CommandsControl basic system functions of the oscilloscope. S

Pagina 575 - :SEARch:RUNT:SOURce

616 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide26 :SYSTem Commands Introduction to:SYSTemCommandsSYSTem subsystem comman

Pagina 576 - :SEARch:RUNT:TIME

:SYSTem Commands 26Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 617 :SYSTem:DATE (see page 860)Command Syntax:SYSTem:DATE

Pagina 577 - :SEARch:TRANsition Commands

618 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide26 :SYSTem Commands :SYSTem:DSP (see page 860)Command Syntax:SYSTem:DSP &

Pagina 578 - :SEARch:TRANsition:QUALifier

:SYSTem Commands 26Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 619 :SYSTem:ERRor (see page 860)Query Syntax:SYSTem:ERRor

Pagina 579 - :SEARch:TRANsition:SLOPe

62 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference *SRE <mask> (see page 144)*SRE? (see page

Pagina 580 - :SEARch:TRANsition:SOURce

620 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide26 :SYSTem Commands :SYSTem:LOCK (see page 860)Command Syntax:SYSTem:LOCK

Pagina 581 - :SEARch:TRANsition:TIME

:SYSTem Commands 26Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 621 :SYSTem:PRESet (see page 860)Command Syntax:SYSTem:PR

Pagina 582 - :SEARch:SERial:CAN Commands

622 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide26 :SYSTem Commands Digital Channel Menu (MSO models only)Channel 0 - 7 O

Pagina 583 - :SEARch Commands 25

:SYSTem Commands 26Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 623 See Also • "Introduction to Common (*) Commands&

Pagina 584

624 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide26 :SYSTem Commands :SYSTem:PROTection:LOCK (see page 860)Command Syntax:

Pagina 585 - :SEARch:SERial:CAN:PATTern:ID

:SYSTem Commands 26Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 625 :SYSTem:SETup (see page 860)Command Syntax:SYSTem:SET

Pagina 586

626 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide26 :SYSTem Commands ' Write setup string back to oscilloscope using

Pagina 587 - :SEARch:SERial:CAN:MODE

:SYSTem Commands 26Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 627 :SYSTem:TIME (see page 860)Command Syntax:SYSTem:TIME

Pagina 588 - :SEARch:SERial:I2S Commands

628 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide26 :SYSTem Commands

Pagina 589 - :SEARch:SERial:I2S:AUDio

629 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide27:TIMebase CommandsControl all horizontal sweep functions. See "I

Pagina 590 - :SEARch:SERial:I2S:MODE

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 63 Table 3 Root (:) Commands SummaryCommand Query

Pagina 591

630 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide27 :TIMebase Commands Introduction to:TIMebaseCommandsThe TIMebase subsys

Pagina 592

:TIMebase Commands 27Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 631 :TIMebase:MODE (see page 860)Command Syntax:TIMebas

Pagina 593 - :SEARch:SERial:I2S:RANGe

632 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide27 :TIMebase Commands :TIMebase:POSition (see page 860)Command Syntax:TIM

Pagina 594 - :SEARch:SERial:IIC Commands

:TIMebase Commands 27Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 633 :TIMebase:RANGe (see page 860)Command Syntax:TIMeba

Pagina 595 - :SEARch:SERial:IIC:MODE

634 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide27 :TIMebase Commands :TIMebase:REFerence (see page 860)Command Syntax:TI

Pagina 596

:TIMebase Commands 27Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 635 :TIMebase:SCALe (see page 860)Command Syntax:TIMeba

Pagina 597

636 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide27 :TIMebase Commands :TIMebase:VERNier (see page 860)Command Syntax:TIMe

Pagina 598

:TIMebase Commands 27Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 637 :TIMebase:WINDow:POSition (see page 860)Command Syn

Pagina 599

638 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide27 :TIMebase Commands :TIMebase:WINDow:RANGe (see page 860)Command Syntax

Pagina 600 - :SEARch:SERial:IIC:QUALifier

:TIMebase Commands 27Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 639 :TIMebase:WINDow:SCALe (see page 860)Command Syntax

Pagina 601 - :SEARch:SERial:LIN Commands

64 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :DIGitize [<source>[,..,<source>]]

Pagina 602 - :SEARch:SERial:LIN:ID

640 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide27 :TIMebase Commands

Pagina 603 - :SEARch:SERial:LIN:MODE

641 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28:TRIGger CommandsControl the trigger modes and parameters for each tr

Pagina 604

642 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands •Edge triggering— identifies a trigger by looking fo

Pagina 605

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 643 General :TRIGger CommandsTable 104 General :TRIGger

Pagina 606

644 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:HFReject (see page 860)Command Syntax:TRIGg

Pagina 607 - :SEARch:SERial:SPI Commands

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 645 :TRIGger:HOLDoff (see page 860)Command Syntax:TRIGge

Pagina 608 - :SEARch:SERial:SPI:MODE

646 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:LEVel:HIGH (see page 860)Command Syntax:TRI

Pagina 609

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 647 :TRIGger:LEVel:LOW (see page 860)Command Syntax:TRIG

Pagina 610

648 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:MODE (see page 860)Command Syntax:TRIGger:M

Pagina 611 - :SEARch:SERial:UART Commands

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 649 :TRIGger:NREJect (see page 860)Command Syntax:TRIGge

Pagina 612 - :SEARch:SERial:UART:DATA

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 65 :PRINt [<options>] (see page 179)n/a <

Pagina 613 - :SEARch:SERial:UART:MODE

650 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:SWEep (see page 860)Command Syntax:TRIGger:

Pagina 614 - :SEARch:SERial:UART:QUALifier

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 651 :TRIGger:EBURst CommandsThe :TRIGger:EBURst:SOURce c

Pagina 615 - :SYSTem Commands

652 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:EBURst:COUNt (see page 860)Command Syntax:T

Pagina 616

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 653 :TRIGger:EBURst:IDLE (see page 860)Command Syntax:TR

Pagina 617 - :SYSTem:DATE

654 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:EBURst:SLOPe (see page 860)Command Syntax:T

Pagina 618 - :SYSTem:DSP

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 655 :TRIGger:EBURst:SOURce (see page 860)Command Syntax:

Pagina 619 - :SYSTem:ERRor

656 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger[:EDGE] CommandsTable 106 :TRIGger[:EDGE] Co

Pagina 620 - :SYSTem:LOCK

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 657 :TRIGger[:EDGE]:COUPling (see page 860)Command Synta

Pagina 621 - :SYSTem:PRESet

658 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger[:EDGE]:LEVel (see page 860)Command Syntax:T

Pagina 622

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 659 :TRIGger[:EDGE]:REJect (see page 860)Command Syntax:

Pagina 623

66 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference Table 4 :ACQuire Commands SummaryCommand Query

Pagina 624 - :SYSTem:PROTection:LOCK

660 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger[:EDGE]:SLOPe (see page 860)Command Syntax:T

Pagina 625 - :SYSTem:SETup

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 661 :TRIGger[:EDGE]:SOURce (see page 860)Command Syntax:

Pagina 626

662 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:GLITch CommandsTable 107 :TRIGger:GLITch Co

Pagina 627 - :SYSTem:TIME

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 663 :TRIGger:GLITch:RANGe <less_than_time>[suffix]

Pagina 628

664 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:GLITch:GREaterthan (see page 860)Command Sy

Pagina 629 - :TIMebase Commands

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 665 :TRIGger:GLITch:LESSthan (see page 860)Command Synta

Pagina 630

666 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:GLITch:LEVel (see page 860)Command Syntax:T

Pagina 631 - :TIMebase:MODE

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 667 :TRIGger:GLITch:POLarity (see page 860)Command Synta

Pagina 632 - :TIMebase:POSition

668 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:GLITch:QUALifier (see page 860)Command Synt

Pagina 633 - :TIMebase:RANGe

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 669 :TRIGger:GLITch:RANGe (see page 860)Command Syntax:T

Pagina 634 - :TIMebase:REFerence

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 67 :BUS<n>:CLEar (see page 206)n/a <n>

Pagina 635 - :TIMebase:SCALe

670 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:GLITch:SOURce (see page 860)Command Syntax:

Pagina 636 - :TIMebase:VERNier

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 671 :TRIGger:PATTern CommandsTable 108 :TRIGger:PATTern

Pagina 637 - :TIMebase:WINDow:POSition

672 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:PATTern (see page 860)Command Syntax:TRIGge

Pagina 638 - :TIMebase:WINDow:RANGe

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 673 You can only specify an edge on one channel. When an

Pagina 639 - :TIMebase:WINDow:SCALe

674 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:PATTern:FORMat (see page 860)Command Syntax

Pagina 640

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 675 :TRIGger:PATTern:GREaterthan (see page 860)Command S

Pagina 641 - :TRIGger Commands

676 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:PATTern:LESSthan (see page 860)Command Synt

Pagina 642

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 677 :TRIGger:PATTern:QUALifier (see page 860)Command Syn

Pagina 643 - General :TRIGger Commands

678 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands • ":TRIGger:PATTern:LESSthan" on page 676

Pagina 644 - :TRIGger:HFReject

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 679 :TRIGger:PATTern:RANGe (see page 860)Command Syntax:

Pagina 645 - :TRIGger:HOLDoff

68 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference n/a :CALibrate:STATus? (see page 218)<return

Pagina 646 - :TRIGger:LEVel:HIGH

680 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:RUNT CommandsTable 109 :TRIGger:RUNT Comman

Pagina 647 - :TRIGger:LEVel:LOW

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 681 :TRIGger:RUNT:POLarity (see page 860)Command Syntax:

Pagina 648 - :TRIGger:MODE

682 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:RUNT:QUALifier (see page 860)Command Syntax

Pagina 649 - :TRIGger:NREJect

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 683 :TRIGger:RUNT:SOURce (see page 860)Command Syntax:TR

Pagina 650 - :TRIGger:SWEep

684 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:RUNT:TIME (see page 860)Command Syntax:TRIG

Pagina 651 - :TRIGger:EBURst Commands

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 685 :TRIGger:SHOLd CommandsTable 110 :TRIGger:SHOLd Comm

Pagina 652 - :TRIGger:EBURst:COUNt

686 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:SHOLd:SLOPe (see page 860)Command Syntax:TR

Pagina 653 - :TRIGger:EBURst:IDLE

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 687 :TRIGger:SHOLd:SOURce:CLOCk (see page 860)Command Sy

Pagina 654 - :TRIGger:EBURst:SLOPe

688 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:SHOLd:SOURce:DATA (see page 860)Command Syn

Pagina 655 - :TRIGger:EBURst:SOURce

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 689 :TRIGger:SHOLd:TIME:HOLD (see page 860)Command Synta

Pagina 656 - :TRIGger[:EDGE] Commands

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 69 :CHANnel<n>:PROBe <attenuation> (se

Pagina 657 - :TRIGger[:EDGE]:COUPling

690 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:SHOLd:TIME:SETup (see page 860)Command Synt

Pagina 658 - :TRIGger[:EDGE]:LEVel

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 691 :TRIGger:TRANsition CommandsThe :TRIGger:TRANsition

Pagina 659 - :TRIGger[:EDGE]:REJect

692 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:TRANsition:QUALifier (see page 860)Command

Pagina 660 - :TRIGger[:EDGE]:SLOPe

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 693 :TRIGger:TRANsition:SLOPe (see page 860)Command Synt

Pagina 661 - :TRIGger[:EDGE]:SOURce

694 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:TRANsition:SOURce (see page 860)Command Syn

Pagina 662 - :TRIGger:GLITch Commands

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 695 :TRIGger:TRANsition:TIME (see page 860)Command Synta

Pagina 663

696 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:TV CommandsTable 112 :TRIGger:TV Commands S

Pagina 664 - :TRIGger:GLITch:GREaterthan

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 697 :TRIGger:TV:LINE (see page 860)Command Syntax:TRIGge

Pagina 665 - :TRIGger:GLITch:LESSthan

698 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:TV:MODE (see page 860)Command Syntax:TRIGge

Pagina 666 - :TRIGger:GLITch:LEVel

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 699 :TRIGger:TV:POLarity (see page 860)Command Syntax:TR

Pagina 667 - :TRIGger:GLITch:POLarity

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 7 ContentsIn This Book 31 What's NewWhat's New in Version 1.10 26V

Pagina 668 - :TRIGger:GLITch:QUALifier

70 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference Table 8 :DEMO Commands SummaryCommand Query Opt

Pagina 669 - :TRIGger:GLITch:RANGe

700 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:TV:SOURce (see page 860)Command Syntax:TRIG

Pagina 670 - :TRIGger:GLITch:SOURce

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 701 :TRIGger:TV:STANdard (see page 860)Command Syntax:TR

Pagina 671 - :TRIGger:PATTern Commands

702 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:USB CommandsTable 114 :TRIGger:USB Commands

Pagina 672 - :TRIGger:PATTern

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 703 :TRIGger:USB:SOURce:DMINus (see page 860)Command Syn

Pagina 673 - :TRIGger Commands 28

704 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:USB:SOURce:DPLus (see page 860)Command Synt

Pagina 674 - :TRIGger:PATTern:FORMat

:TRIGger Commands 28Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 705 :TRIGger:USB:SPEed (see page 860)Command Syntax:TRIG

Pagina 675 - :TRIGger:PATTern:GREaterthan

706 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide28 :TRIGger Commands :TRIGger:USB:TRIGger (see page 860)Command Syntax:TR

Pagina 676 - :TRIGger:PATTern:LESSthan

707 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29:WAVeform CommandsProvide access to waveform data. See "Introduc

Pagina 677 - :TRIGger:PATTern:QUALifier

708 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:POINts:MODE <points_mode> (see page

Pagina 678

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 709 Introduction to:WAVeformCommandsThe WAVeform subsys

Pagina 679 - :TRIGger:PATTern:RANGe

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 71 Table 10 :DISPlay Commands SummaryCommand Query

Pagina 680 - :TRIGger:RUNT Commands

710 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands Data Acquisition TypesThere are four types of wavef

Pagina 681 - :TRIGger:RUNT:POLarity

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 711 Normal data consists of the last data point (hit) i

Pagina 682 - :TRIGger:RUNT:QUALifier

712 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands The high resolution (smoothing) mode is used to red

Pagina 683 - :TRIGger:RUNT:SOURce

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 713 Use the :WAVeform:UNSigned command (see page 735) t

Pagina 684 - :TRIGger:RUNT:TIME

714 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands Digital Channel Data (MSO models only)The waveform

Pagina 685 - :TRIGger:SHOLd Commands

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 715 :WAVeform:BYTeorder (see page 860)Command Syntax:WA

Pagina 686 - :TRIGger:SHOLd:SLOPe

716 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:COUNt (see page 860)Query Syntax:WAVeform

Pagina 687 - :TRIGger:SHOLd:SOURce:CLOCk

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 717 :WAVeform:DATA (see page 860)Query Syntax:WAVeform:

Pagina 688 - :TRIGger:SHOLd:SOURce:DATA

718 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands '' <header><waveform_data><

Pagina 689 - :TRIGger:SHOLd:TIME:HOLD

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 719 :WAVeform:FORMat (see page 860)Command Syntax:WAVef

Pagina 690 - :TRIGger:SHOLd:TIME:SETup

72 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference Table 12 :FUNCtion Commands SummaryCommand Quer

Pagina 691 - :TRIGger:TRANsition Commands

720 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:POINts (see page 860)Command Syntax:WAVef

Pagina 692 - :TRIGger:TRANsition:QUALifier

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 721 | 4000000 | 8000000 | <maximum # points>}if w

Pagina 693 - :TRIGger:TRANsition:SLOPe

722 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:POINts:MODE (see page 860)Command Syntax:

Pagina 694 - :TRIGger:TRANsition:SOURce

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 723 The :WAVeform:POINts:MODE? query returns the curren

Pagina 695 - :TRIGger:TRANsition:TIME

724 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:PREamble (see page 860)Query Syntax:WAVef

Pagina 696 - :TRIGger:TV Commands

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 725 • ":DIGitize" on page 163• ":WAVefo

Pagina 697 - :TRIGger:TV:LINE

726 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands sngYIncrement = Preamble(7)sngYOrigin = Preamble(8)

Pagina 698 - :TRIGger:TV:MODE

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 727 :WAVeform:SEGMented:COUNt (see page 860)Query Synta

Pagina 699 - :TRIGger:TV:POLarity

728 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:SEGMented:TTAG (see page 860)Query Syntax

Pagina 700 - :TRIGger:TV:SOURce

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 729 :WAVeform:SOURce (see page 860)Command Syntax:WAVef

Pagina 701 - :TRIGger:TV:STANdard

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 73 :FUNCtion:REFerence <level> (see page 280

Pagina 702 - :TRIGger:USB Commands

730 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands Return Format <source><NL><source>

Pagina 703 - :TRIGger:USB:SOURce:DMINus

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 731 ' where <preamble_block> is:' FORMA

Pagina 704 - :TRIGger:USB:SOURce:DPLus

732 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands " V" + vbCrLfstrOutput = strOutput + &quo

Pagina 705 - :TRIGger:USB:SPEed

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 733 :WAVeform:SOURce:SUBSource (see page 860)Command Sy

Pagina 706 - :TRIGger:USB:TRIGger

734 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:TYPE (see page 860)Query Syntax:WAVeform:

Pagina 707 - :WAVeform Commands

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 735 :WAVeform:UNSigned (see page 860)Command Syntax:WAV

Pagina 708

736 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:VIEW (see page 860)Command Syntax:WAVefor

Pagina 709

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 737 :WAVeform:XINCrement (see page 860)Query Syntax:WAV

Pagina 710

738 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:XORigin (see page 860)Query Syntax:WAVefo

Pagina 711 - :WAVeform Commands 29

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 739 :WAVeform:XREFerence (see page 860)Query Syntax:WAV

Pagina 712

74 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :HARDcopy:FFEed {{0 | OFF} | {1 | ON}} (see pag

Pagina 713

740 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:YINCrement (see page 860)Query Syntax:WAV

Pagina 714

:WAVeform Commands 29Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 741 :WAVeform:YORigin (see page 860)Query Syntax:WAVefo

Pagina 715 - :WAVeform:BYTeorder

742 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide29 :WAVeform Commands :WAVeform:YREFerence (see page 860)Query Syntax:WAV

Pagina 716 - :WAVeform:COUNt

743 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide30:WGEN CommandsWhen the built- in waveform generator is licensed (Opti

Pagina 717 - :WAVeform:DATA

744 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide30 :WGEN Commands Introduction to:WGENCommandsThe :WGEN subsystem provide

Pagina 718

:WGEN Commands 30Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 745 :WGEN:FREQuency (see page 860)Command Syntax:WGEN:FREQu

Pagina 719 - :WAVeform:FORMat

746 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide30 :WGEN Commands :WGEN:FUNCtion (see page 860)Command Syntax:WGEN:FUNCti

Pagina 720 - :WAVeform:POINts

:WGEN Commands 30Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 747 For all waveform types, the output amplitude, into 50 

Pagina 721

748 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide30 :WGEN Commands :WGEN:FUNCtion:PULSe:WIDTh (see page 860)Command Syntax

Pagina 722 - :WAVeform:POINts:MODE

:WGEN Commands 30Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 749 :WGEN:FUNCtion:RAMP:SYMMetry (see page 860)Command Synt

Pagina 723

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 75 Table 14 :LISTer Commands SummaryCommand Query

Pagina 724 - #Vsteps =

750 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide30 :WGEN Commands :WGEN:FUNCtion:SQUare:DCYCle (see page 860)Command Synt

Pagina 725

:WGEN Commands 30Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 751 :WGEN:OUTPut (see page 860)Command Syntax:WGEN:OUTPut &

Pagina 726

752 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide30 :WGEN Commands :WGEN:OUTPut:LOAD (see page 860)Command Syntax:WGEN:OUT

Pagina 727 - :WAVeform:SEGMented:COUNt

:WGEN Commands 30Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 753 :WGEN:PERiod (see page 860)Command Syntax:WGEN:PERiod &

Pagina 728 - :WAVeform:SEGMented:TTAG

754 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide30 :WGEN Commands :WGEN:RST (see page 860)Command Syntax:WGEN:RSTThe :WGE

Pagina 729 - :WAVeform:SOURce

:WGEN Commands 30Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 755 :WGEN:VOLTage (see page 860)Command Syntax:WGEN:VOLTage

Pagina 730

756 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide30 :WGEN Commands :WGEN:VOLTage:HIGH (see page 860)Command Syntax:WGEN:VO

Pagina 731

:WGEN Commands 30Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 757 :WGEN:VOLTage:LOW (see page 860)Command Syntax:WGEN:VOL

Pagina 732

758 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide30 :WGEN Commands :WGEN:VOLTage:OFFSet (see page 860)Command Syntax:WGEN:

Pagina 733 - :WAVeform:SOURce:SUBSource

759 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide31:WMEMory<r> CommandsControl reference waveforms.Table 117 :WMEM

Pagina 734 - :WAVeform:TYPE

76 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference n/a :MARKer:XDELta? (see page 316)<return_va

Pagina 735 - :WAVeform:UNSigned

760 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide31 :WMEMory<r> Commands :WMEMory<r>:YRANge <range>[suff

Pagina 736 - :WAVeform:VIEW

:WMEMory<r> Commands 31Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 761 :WMEMory<r>:CLEar (see page 860)Comma

Pagina 737 - :WAVeform:XINCrement

762 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide31 :WMEMory<r> Commands :WMEMory<r>:DISPlay (see page 860)Com

Pagina 738 - :WAVeform:XORigin

:WMEMory<r> Commands 31Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 763 :WMEMory<r>:LABel (see page 860)Comma

Pagina 739 - :WAVeform:XREFerence

764 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide31 :WMEMory<r> Commands :WMEMory<r>:SAVE (see page 860)Comman

Pagina 740 - :WAVeform:YINCrement

:WMEMory<r> Commands 31Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 765 :WMEMory<r>:SKEW (see page 860)Comman

Pagina 741 - :WAVeform:YORigin

766 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide31 :WMEMory<r> Commands :WMEMory<r>:YOFFset (see page 860)Com

Pagina 742 - :WAVeform:YREFerence

:WMEMory<r> Commands 31Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 767 :WMEMory<r>:YRANge (see page 860)Comm

Pagina 743 - :WGEN Commands

768 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide31 :WMEMory<r> Commands :WMEMory<r>:YSCale (see page 860)Comm

Pagina 744

769 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32Obsolete and Discontinued CommandsObsolete commands are older forms o

Pagina 745 - :WGEN:FREQuency

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 77 :MEASure:COUNter [<source>] (see page 336

Pagina 746 - :WGEN:FUNCtion

770 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :CHANnel<n>:INPut (see page 7

Pagina 747 - <signal><NL>

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 771 :MEASure:TMAX (see page 795):MEASur

Pagina 748 - :WGEN:FUNCtion:PULSe:WIDTh

772 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands DiscontinuedCommandsDiscontinued co

Pagina 749 - :WGEN:FUNCtion:RAMP:SYMMetry

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 773 DiscontinuedParametersSome previous

Pagina 750 - :WGEN:FUNCtion:SQUare:DCYCle

774 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :CHANnel:ACTivity (see page 860)Com

Pagina 751 - :WGEN:OUTPut

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 775 :CHANnel:LABel (see page 860)Comman

Pagina 752 - :WGEN:OUTPut:LOAD

776 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :CHANnel:THReshold (see page 860)Co

Pagina 753 - :WGEN:PERiod

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 777 :CHANnel2:SKEW (see page 860)Comman

Pagina 754 - :WGEN:RST

778 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :CHANnel<n>:INPut (see page 8

Pagina 755 - the same as a high- level

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 779 :CHANnel<n>:PMODe (see page 8

Pagina 756 - :WGEN:VOLTage:HIGH

78 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :MEASure:DUTYcycle [<source>] (see page 3

Pagina 757 - :WGEN:VOLTage:LOW

780 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :DISPlay:CONNect (see page 860)Comm

Pagina 758 - :WGEN:VOLTage:OFFSet

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 781 :DISPlay:ORDer (see page 860)Query

Pagina 759 - :WMEMory<r> Commands

782 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :ERASe (see page 860)Command Syntax

Pagina 760

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 783 :EXTernal:PMODe (see page 860)Comma

Pagina 761 - :WMEMory<r>:CLEar

784 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :FUNCtion:SOURce (see page 860)Comm

Pagina 762 - :WMEMory<r>:DISPlay

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 785 :FUNCtion:VIEW (see page 860)Comman

Pagina 763 - :WMEMory<r>:LABel

786 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :HARDcopy:DESTination (see page 860

Pagina 764 - :WMEMory<r>:SAVE

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 787 :HARDcopy:FILename (see page 860)Co

Pagina 765 - :WMEMory<r>:SKEW

788 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :HARDcopy:GRAYscale (see page 860)C

Pagina 766 - :WMEMory<r>:YOFFset

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 789 :HARDcopy:IGColors (see page 860)Co

Pagina 767 - :WMEMory<r>:YRANge

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 79 :MEASure:NEDGes [<source>] (see page 345)

Pagina 768 - :WMEMory<r>:YSCale

790 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :HARDcopy:PDRiver (see page 860)Com

Pagina 769 - Commands" on page 860)

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 791 :MEASure:LOWer (see page 860)Comman

Pagina 770

792 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :MEASure:SCRatch (see page 860)Comm

Pagina 771

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 793 :MEASure:TDELta (see page 860)Query

Pagina 772

794 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :MEASure:THResholds (see page 860)C

Pagina 773 - Parameters

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 795 :MEASure:TMAX (see page 860)Command

Pagina 774 - :CHANnel:ACTivity

796 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :MEASure:TMIN (see page 860)Command

Pagina 775 - :CHANnel:LABel

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 797 :MEASure:TSTArt (see page 860)Comma

Pagina 776 - :CHANnel:THReshold

798 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :MEASure:TSTOp (see page 860)Comman

Pagina 777 - :CHANnel2:SKEW

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 799 :MEASure:TVOLt (see page 860)Query

Pagina 778 - :CHANnel<n>:INPut

8 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide Other Ways of Sending Commands 55Telnet Sockets 55Sending SCPI Commands Us

Pagina 779 - :CHANnel<n>:PMODe

80 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :MEASure:PERiod [<source>] (see page 351)

Pagina 780 - :DISPlay:CONNect

800 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands <value> ::= time in seconds o

Pagina 781 - :DISPlay:ORDer

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 801 :MEASure:UPPer (see page 860)Comman

Pagina 782 - Command Syntax

802 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :MEASure:VDELta (see page 860)Query

Pagina 783 - :EXTernal:PMODe

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 803 :MEASure:VSTArt (see page 860)Comma

Pagina 784 - :FUNCtion:SOURce

804 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :MEASure:VSTOp (see page 860)Comman

Pagina 785 - :FUNCtion:VIEW

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 805 :MTESt:AMASk:{SAVE | STORe} (see pa

Pagina 786 - :HARDcopy:DESTination

806 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :MTESt:AVERage (see page 860)Comman

Pagina 787 - :HARDcopy:FILename

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 807 :MTESt:AVERage:COUNt (see page 860)

Pagina 788 - :HARDcopy:GRAYscale

808 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :MTESt:LOAD (see page 860)Command S

Pagina 789 - :HARDcopy:IGColors

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 809 :MTESt:RUMode (see page 860)Command

Pagina 790 - :HARDcopy:PDRiver

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 81 :MEASure:PWIDth [<source>] (see page 355)

Pagina 791 - :MEASure:LOWer

810 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :MTESt:RUMode:SOFailure (see page 8

Pagina 792 - :MEASure:SCRatch

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 811 :MTESt:{STARt | STOP} (see page 860

Pagina 793 - :MEASure:TDELta

812 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :MTESt:TRIGger:SOURce (see page 860

Pagina 794 - :MEASure:THResholds

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 813 :PRINt? (see page 860)Query Syntax:

Pagina 795 - :MEASure:TMAX

814 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands See Also • "Introduction to Ro

Pagina 796 - :MEASure:TMIN

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 815 :SAVE:IMAGe:AREA (see page 860)Quer

Pagina 797 - :MEASure:TSTArt

816 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :SBUS<n>:LIN:SIGNal:DEFinitio

Pagina 798 - :MEASure:TSTOp

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 817 :TIMebase:DELay (see page 860)Comma

Pagina 799 - :MEASure:TVOLt

818 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands :TRIGger:THReshold (see page 860)Co

Pagina 800

Obsolete and Discontinued Commands 32Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 819 :TRIGger:TV:TVMode (see page 860)Co

Pagina 801 - :MEASure:UPPer

82 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :MEASure:SOURce <source1> [,<source2&g

Pagina 802 - :MEASure:VDELta

820 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide32 Obsolete and Discontinued Commands

Pagina 803 - :MEASure:VSTArt

821 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide33Error Messages-440, Query UNTERMINATED after indefinite response-430,

Pagina 804 - :MEASure:VSTOp

822 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide33 Error Messages -314, Save/recall memory lost-313, Calibration memory l

Pagina 805 - :MTESt:AMASk:{SAVE

Error Messages 33Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 823 -255, Directory full-254, Media full-253, Corrupt media

Pagina 806 - :MTESt:AVERage

824 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide33 Error Messages -223, Too much data-222, Data out of range-221, Setting

Pagina 807 - :MTESt:AVERage:COUNt

Error Messages 33Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 825 -151, Invalid string data-150, String data error-148, C

Pagina 808 - :MTESt:LOAD

826 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide33 Error Messages -112, Program mnemonic too long-109, Missing parameter-

Pagina 809 - :MTESt:RUMode

Error Messages 33Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 827 +103, Write Error+104, Illegal Operation+105, Print Can

Pagina 810 - :MTESt:RUMode:SOFailure

828 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide33 Error Messages

Pagina 811 - :MTESt:{STARt

829 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34Status ReportingStatus Reporting Data Structures 831Status Byte Regis

Pagina 812 - :MTESt:TRIGger:SOURce

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 83 n/a :MEASure:TVALue? <value>, [<slope&

Pagina 813

830 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34 Status Reporting • To monitor an event, first clear the event; then, e

Pagina 814 - Old Print Option: Is Now:

Status Reporting 34Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 831 Status Reporting Data StructuresThe following figure

Pagina 815 - :SAVE:IMAGe:AREA

832 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34 Status Reporting • Table 66• Table 61The status registers picture abov

Pagina 816 - <value> ::= LIN

Status Reporting 34Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 833 Status Byte Register (STB)The Status Byte Register is

Pagina 817 - :TIMebase:DELay

834 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34 Status Reporting The next program prints 0xD1 and clears bit 6 (RQS) a

Pagina 818 - :TRIGger:THReshold

Status Reporting 34Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 835 Service Request Enable Register (SRE)Setting the Serv

Pagina 819 - :TRIGger:TV:TVMode

836 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34 Status Reporting Trigger Event Register (TER)This register sets the TR

Pagina 820

Status Reporting 34Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 837 Output QueueThe output queue stores the oscilloscope-

Pagina 821 - Error Messages

838 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34 Status Reporting Message QueueThe message queue contains the text of t

Pagina 822

Status Reporting 34Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 839 (Standard) Event Status Register (ESR)The (Standard)

Pagina 823 - Error Messages 33

84 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :MEASure:VMAX [<source>] (see page 374):M

Pagina 824

840 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34 Status Reporting (Standard) Event Status Enable Register (ESE)To allow

Pagina 825

Status Reporting 34Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 841 Error QueueAs errors are detected, they are placed in

Pagina 826

842 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34 Status Reporting Operation Status Event Register (:OPERegister[:EVENt]

Pagina 827

Status Reporting 34Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 843 Operation Status Condition Register (:OPERegister:CON

Pagina 828

844 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34 Status Reporting Arm Event Register (AER)This register sets bit 5 (Wai

Pagina 829 - Status Reporting

Status Reporting 34Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 845 Overload Event Register (:OVLRegister)The Overload Ev

Pagina 830

846 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34 Status Reporting Mask Test Event Event Register (:MTERegister[:EVENt])

Pagina 831 - From Mask Test

Status Reporting 34Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 847 Clearing Registers and QueuesThe *CLS common command

Pagina 832

848 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide34 Status Reporting Status Reporting Decision Chartyesno Do you wantto do

Pagina 833 - Status Byte Register (STB)

849 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide35Synchronizing AcquisitionsSynchronization in the Programming Flow 850

Pagina 834

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 85 n/a :MEASure:VTIMe? <vtime>[,<source&g

Pagina 835 - Status Reporting 34

850 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide35 Synchronizing Acquisitions Synchronization in the Programming FlowMost

Pagina 836 - Trigger Event Register (TER)

Synchronizing Acquisitions 35Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 851 Blocking SynchronizationUse the :DIGitize c

Pagina 837 - Output Queue

852 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide35 Synchronizing Acquisitions Polling Synchronization With TimeoutThis ex

Pagina 838 - Message Queue

Synchronizing Acquisitions 35Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 853 myScope.WriteString ":OPERegister:COND

Pagina 839

854 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide35 Synchronizing Acquisitions Synchronizing with a Single-Shot Device Und

Pagina 840

Synchronizing Acquisitions 35Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 855 ' ------------------------------------

Pagina 841 - Error Queue

856 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide35 Synchronizing Acquisitions Synchronization with an Averaging Acquisiti

Pagina 842

Synchronizing Acquisitions 35Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 857 ' Save *ESE (Standard Event Status Ena

Pagina 843 - Status Condition Register

858 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide35 Synchronizing Acquisitions

Pagina 844 - Arm Event Register (AER)

859 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide36More About Oscilloscope CommandsCommand Classifications 860Valid Comm

Pagina 845

86 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference Table 17 :MTESt Commands SummaryCommand Query O

Pagina 846

860 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide36 More About Oscilloscope Commands Command ClassificationsTo help you us

Pagina 847 - Clearing Registers and Queues

More About Oscilloscope Commands 36Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 861 Valid Command/Query Strings• "Pr

Pagina 848

862 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide36 More About Oscilloscope Commands on whether or not you have included t

Pagina 849 - Synchronizing Acquisitions

More About Oscilloscope Commands 36Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 863 In the oscilloscope programmer's

Pagina 850 - Retrieve Results

864 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide36 More About Oscilloscope Commands No space or separator is allowed betw

Pagina 851 - Blocking Synchronization

More About Oscilloscope Commands 36Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 865 All numbers must be strings of ASCII

Pagina 852

866 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide36 More About Oscilloscope Commands :<subsystem>:<function>&l

Pagina 853 - Synchronizing Acquisitions 35

More About Oscilloscope Commands 36Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 867 Query Return ValuesCommand headers im

Pagina 854

868 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide36 More About Oscilloscope Commands All Oscilloscope Commands Are Sequent

Pagina 855

869 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37Programming ExamplesVISA COM Examples 870VISA Examples 897SICL Exampl

Pagina 856

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 87 :MTESt:RMODe:FACTion:MEASure {{0 | OFF} | {1 |

Pagina 857

870 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples VISA COM Examples• "VISA COM Example in Visu

Pagina 858

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 871 ' Create the VISA COM I/O resource.Set myMgr

Pagina 859 - Query Return Values 867

872 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples ' Use auto-scale to automatically configure

Pagina 860 - Command Classifications

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 873 ' Set the acquisition type (NORMal, PEAK, AV

Pagina 861 - Program Data

874 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples DoCommand ":HARDcopy:INKSaver OFF"Dim b

Pagina 862

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 875 intFormat = Preamble(0)intType = Preamble(1)lngPo

Pagina 863 - Common Command Headers

876 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples ' Set up output file:strPath = "c:\scop

Pagina 864 - Program Data Syntax Rules

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 877 Dim strErrors As StringmyScope.WriteIEEEBlock com

Pagina 865 - Duplicate Mnemonics

878 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples myScope.WriteString queryDoQueryNumbers = myScope

Pagina 866

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 879 MsgBox "VISA COM Error: " + vbCrLf + Er

Pagina 867 - Query Return Values

88 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :MTESt:SOURce <source> (see page 417):MTE

Pagina 868

880 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples );myScope.SetTimeoutSeconds(10);// Initialize - s

Pagina 869 - Programming Examples

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 881 // Set trigger mode (EDGE, PULSe, PATTern, etc.,

Pagina 870 - VISA COM Examples

882 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples Console.WriteLine("Acquire type: {0}",m

Pagina 871 - Programming Examples 37

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 883 fStream.Write(ResultsArray, 0, nLength);fStream.C

Pagina 872

884 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples }else if (fType == 3.0){Console.WriteLine("A

Pagina 873

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 885 {private ResourceManagerClass m_ResourceManager;p

Pagina 874

886 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples // Get the result number.double fResult;fResult =

Pagina 875

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 887 strInstrumentError = m_IoObject.ReadString();if (

Pagina 876

888 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples }}}VISA COM Example in Visual Basic .NETTo compil

Pagina 877

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 889 )myScope.SetTimeoutSeconds(10)' Initialize -

Pagina 878

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 89 :RECall:SETup[:STARt] [<file_spec>] (see

Pagina 879 - VISA COM Example in C#

890 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples myScope.DoCommand(":TRIGger:EDGE:LEVel 1.5&q

Pagina 880

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 891 strPath = "c:\scope\config\setup.stp"Da

Pagina 881

892 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples ' Set the waveform points mode.myScope.DoCom

Pagina 882

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 893 Console.WriteLine("Waveform X reference: {0:

Pagina 883

894 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples ' Clear the interface.m_IoObject.IO.Clear()E

Pagina 884

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 895 Public Function DoQueryNumbers(ByVal strQuery As

Pagina 885

896 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples m_IoObject = New FormattedIO488Class()' Open

Pagina 886

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 897 VISA Examples• "VISA Example in C" on p

Pagina 887

898 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples * This program illustrates a few commonly-used pr

Pagina 888

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 899 /* Analyze the captured waveform. */analyze();/*

Pagina 889

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 9 :AUToscale:CHANnels 160:AUToscale:FDEBug 161:BLANk 162:DIGitize 163:MTEena

Pagina 890

90 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :SAVE:MASK[:STARt] [<file_spec>] (see pag

Pagina 891

900 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples printf("Read setup string query (%d bytes).\

Pagina 892

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 901 {double wav_format;double acq_type;double wav_poi

Pagina 893

902 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples /* Set the waveform source. */do_command(":W

Pagina 894

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 903 x_reference = dbl_results[6];printf("Wavefor

Pagina 895

904 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples char message[80];int data_length;strcpy(message,

Pagina 896

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 905 char *query;{char message[80];strcpy(message, que

Pagina 897 - VISA Examples

906 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples err = viQueryf(vi, ":SYSTem:ERRor?\n",

Pagina 898

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 907 ' features of your Agilent oscilloscope.&apo

Pagina 899

908 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples End Sub'' Initialize the oscilloscope t

Pagina 900

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 909 ' Output setup string to a file:Dim strPath

Pagina 901

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 91 Table 21 General :SBUS<n> Commands Summar

Pagina 902

910 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples ' Capture an acquisition using :DIGitize.&ap

Pagina 903

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 911 ' Set the waveform points mode.DoCommand &qu

Pagina 904

912 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples ElseIf intType = 1 ThenDebug.Print "Acquisit

Pagina 905

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 913 ' Close output file.Close hFile ' Close

Pagina 906 - VISA Example in Visual Basic

914 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples err = viVScanf(vi, "%lf" + vbLf, VarPtr

Pagina 907

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 915 If (err <> VI_SUCCESS) Then HandleVISAError

Pagina 908

916 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples MsgBox "*** VISA Error : " + strVisaErr

Pagina 909

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 917 namespace InfiniiVision{class VisaInstrumentApp{p

Pagina 910

918 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples myScope.DoCommand("*CLS");myScope.DoCom

Pagina 911

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 919 // Set horizontal scale and position.myScope.DoCo

Pagina 912

92 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :SBUS<n>:CAN:TRIGger <condition> (s

Pagina 913

920 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples // Download the screen image.// -----------------

Pagina 914

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 921 double fType = fResultsArray[1];if (fType == 0.0)

Pagina 915

922 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples writer.WriteLine("{0:f9}, {1:f6}",fXori

Pagina 916 - VISA Example in C#

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 923 nLength);CheckVisaStatus(nViStatus);// Write the

Pagina 917

924 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples // Return string results.return fResultsArray;}pu

Pagina 918

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 925 {double[] fResultsArray;fResultsArray = new doubl

Pagina 919

926 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples }Console.Write(strInstrumentError);}} while (!str

Pagina 920

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 927 VISA Example in Visual Basic .NETTo compile and r

Pagina 921

928 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples New VisaInstrument("USB0::0x0957::0x17A6::US

Pagina 922

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 929 myScope.DoQueryString(":TRIGger:EDGE:SOURce?

Pagina 923

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 93 :SBUS<n>:I2S:SOURce:CLOCk <source>

Pagina 924

930 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples ' Read setup string from file.strPath = &quo

Pagina 925

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 931 ' ------------------------------------------

Pagina 926

932 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples Dim fXreference As Double = fResultsArray(6)Conso

Pagina 927

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 933 ' Open a VISA resource session.OpenSession()

Pagina 928

934 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples ' Return string results.Return strResultsEnd

Pagina 929

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 935 Dim strWithNewline As StringstrWithNewline = [Str

Pagina 930

936 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples CheckVisaStatus(nViStatus)' Write and read b

Pagina 931

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 937 If nViStatus < visa32.VI_SUCCESS ThenDim strEr

Pagina 932

938 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples SICL Examples• "SICL Example in C" on p

Pagina 933

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 939 #include <stdio.h> /* For printf(). */#incl

Pagina 934

94 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :SBUS<n>:I2S:TRIGger:RANGe <lower>,

Pagina 935

940 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples /* Capture data. */capture();/* Analyze the captu

Pagina 936

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 941 do_command(":TRIGger:EDGE:LEVel 1.5");d

Pagina 937

942 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples fclose (fp);printf("Read setup string (%d by

Pagina 938 - SICL Examples

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 943 fp);fclose (fp);printf("Wrote screen image (

Pagina 939

944 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples }else if (acq_type == 3.0){printf("Acquire t

Pagina 940

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 945 {char message[80];strcpy(message, command);strcat

Pagina 941

946 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples check_instrument_errors();}/* Query for numbers r

Pagina 942

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 947 strcat(str_out, str_err_val);ipromptf(id, ":

Pagina 943

948 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples ' For Sleep subroutine.Private Declare Sub S

Pagina 944

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 949 ErrorHandler:MsgBox "*** Error :"+Error

Pagina 945

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 95 :SBUS<n>:IIC:TRIGger:PATTern:ADDRess <

Pagina 946

950 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples Put hFile, , byteArray(lngI) ' Write data.Ne

Pagina 947 - SICL Example in Visual Basic

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 951 '' Analyze the captured waveform.'

Pagina 948

952 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples ' Get the number of waveform points availabl

Pagina 949

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 953 Debug.Print "Acquisition type: HRESolution&q

Pagina 950

954 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples MsgBox "Waveform format BYTE data written to

Pagina 951

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 955 On Error GoTo ErrorHandlerDim strResult As String

Pagina 952

956 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples MsgBox "*** Error :"+Error, vbExclamati

Pagina 953

Programming Examples 37Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 957 If Not strOut = "" ThenMsgBox strOut, v

Pagina 954

958 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide37 Programming Examples

Pagina 955

Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 959 IndexSymbols+9.9E+37, infinity representation, 867+9.9E+37, measurement

Pagina 956

96 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :SBUS<n>:LIN:SYNCbreak <value> (see

Pagina 957

960 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's GuideIndex CAL PROTECT button, 216CAL PROTECT switch, 211calculating preshoot

Pagina 958

IndexAgilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 961 cursor mode, 311cursor position, 312, 314, 316, 317, 319cursor read

Pagina 959 - Numerics

962 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's GuideIndex example code, :ACQuire:COMPlete, 189example code, :ACQuire:SEGMente

Pagina 960

IndexAgilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 963 II2C demo signal, 243I2S alignment, 477I2S audio channel, 486I2S cl

Pagina 961

964 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's GuideIndex logic level activity, 774long form, 862low frequency sine with glit

Pagina 962

IndexAgilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 965 multiply math function as g(t) source, 274Nname channels, 229name l

Pagina 963 - , 277, 280

966 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's GuideIndex preamble data, 724preamble metadata, 709predefined logic threshold,

Pagina 964

IndexAgilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 967 runt search qualifier, 574runt search source, 575runt search, pulse

Pagina 965 - function, 277, 280

968 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's GuideIndex square wave duty cycle, waveform generator, 750square waveform gene

Pagina 966

IndexAgilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 969 trigger type, SPI, 533trigger type, UART, 555TRIGger UART commands,

Pagina 967

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 97 :SBUS<n>:SPI:CLOCk:TIMeout <time_value

Pagina 968

970 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's GuideIndex vertical axis range for channels, 237vertical offset for channels,

Pagina 969

98 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide4 Commands Quick Reference :SBUS<n>:SPI:SOURce:MOSI <source> (

Pagina 970

Commands Quick Reference 4Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 99 :SBUS<n>:UART:BITorder <bitorder> (

Commenti su questo manuale