:SAVE Commands 23
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 437
:SAVE:IMAGe[:STARt]
(see page 860)
Command Syntax
:SAVE:IMAGe[:STARt] [<file_name>]
<file_name> ::= quoted ASCII string
The :SAVE:IMAGe[:STARt] command saves an image.
See Also "Introduction to :SAVE Commands" on page 434
":SAVE:IMAGe:FACTors" on page 438
":SAVE:IMAGe:FORMat" on page 439
":SAVE:IMAGe:INKSaver" on page 440
":SAVE:IMAGe:PALette" on page 441
":SAVE:FILename" on page 436
NOTE
Be sure to set the :SAVE:IMAGe:FORMat before saving an image. If the format is NONE, the
save image command will not succeed.
NOTE
If a file extension is provided as part of a specified <file_name>, and it does not match the
extension expected by the format specified in :SAVE:IMAGe:FORMat, the format will be
changed if the extension is a valid image file extension.
NOTE
If the extension ".bmp" is used and the current :SAVE:IMAGe:FORMat is not BMP or BMP8,
the format will be changed to BMP.