Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 751
:WGEN:OUTPut
(see page 860)
Command Syntax
:WGEN:OUTPut <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}
The :WGEN:OUTPut command specifies whether the waveform generator
signal output is ON (1) or OFF (0).
Query Syntax
:WGEN:OUTPut?
The :WGEN:OUTPut? query returns the current state of the waveform
generator output setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :WGEN Commands" on page 744