Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manuale di Servizio Pagina 512

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 970
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 511
512 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:LIN:TRIGger
(see page 860)
Command Syntax
:SBUS<n>:LIN:TRIGger <condition>
<condition> ::= {SYNCbreak | ID | DATA}
The :SBUS<n>:LIN:TRIGger command sets the LIN trigger condition to be:
SYNCbreak — Sync Break.
ID — Frame ID.
Use the :SBUS<n>:LIN:ID command to specify the frame ID.
DATA Frame ID and Data.
Use the :SBUS<n>:LIN:ID command to specify the frame ID.
Use the :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth and
:SBUS<n>:LIN:TRIGger:PATTern:DATA commands to specify the data
string length and value.
Query Syntax :SBUS<n>:LIN:TRIGger?
The :SBUS<n>:LIN:TRIGger? query returns the current LIN trigger value.
Return Format
<condition><NL>
<condition> ::= {SYNC | ID | DATA}
Errors "- 241, Hardware missing" on page 823
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:MODE" on page 648
":SBUS<n>:LIN:TRIGger:ID" on page 513
":SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth" on page 516
":SBUS<n>:LIN:TRIGger:PATTern:DATA" on page 514
":SBUS<n>:LIN:SIGNal:DEFinition" on page 816
":SBUS<n>:LIN:SOURce" on page 509
Vedere la pagina 511
1 2 ... 507 508 509 510 511 512 513 514 515 516 517 ... 969 970

Commenti su questo manuale

Nessun commento