:TRIGger Commands 28
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 675
:TRIGger:PATTern:GREaterthan
(see page 860)
Command Syntax
:TRIGger:PATTern:GREaterthan <greater_than_time>[<suffix>]
<greater_than_time> ::= minimum trigger duration in seconds
in NR3 format
<suffix> ::= {s | ms | us | ns | ps }
The :TRIGger:PATTern:GREaterthan command sets the minimum duration
for the defined pattern when :TRIGger:PATTern:QUALifier is set to
GREaterthan. The command also sets the timeout value when the
:TRIGger:PATTern:QUALifier is set to TIMeout.
Query Syntax
:TRIGger:PATTern:GREaterthan?
The :TRIGger:PATTern:GREaterthan? query returns the minimum duration
time for the defined pattern.
Return Format
<greater_than_time><NL>
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:PATTern" on page 672
":TRIGger:PATTern:QUALifier" on page 677
":TRIGger:MODE" on page 648