756 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
30 :WGEN Commands
:WGEN:VOLTage:HIGH
(see page 860)
Command Syntax
:WGEN:VOLTage:HIGH <high>
<high> ::= high-level voltage in volts, in NR3 format
For Sine, Square, Ramp, Pulse, and Noise waveforms, the
:WGEN:VOLTage:HIGH command specifies the waveform's high- level
voltage. Use the :WGEN:VOLTage:LOW command to specify the low-level
voltage.
You can also specify the high- level and low- level voltages indirectly using
the :WGEN:VOLTage and :WGEN:VOLTage:OFFSet commands. For example,
a high- level voltage of 4 V and a low- level voltage of - 1 V is the same as
an amplitude of 5 V and an offset of 1 V.
Query Syntax
:WGEN:VOLTage:HIGH?
The :WGEN:VOLTage:HIGH? query returns the currently specified waveform
high- level voltage.
Return Format
<high><NL>
<high> ::= high-level voltage in volts, in NR3 format
See Also • "Introduction to :WGEN Commands" on page 744
• ":WGEN:FUNCtion" on page 746
• ":WGEN:VOLTage:LOW" on page 757
• ":WGEN:VOLTage" on page 755
• ":WGEN:VOLTage:OFFSet" on page 758