Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manuale di Servizio Pagina 531

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 970
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 530
:SBUS<n> Commands 24
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 531
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA
(see page 860)
Command Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA <string>
<string> ::= "nn...n" where n ::= {0 |1|X|$}
<string ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $}
The :SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA command defines the SPI
data pattern resource according to the string parameter. This pattern,
along with the data width, control the data pattern searched for in the
data stream.
If the string parameter starts with "0x", it is a hexadecimal string made up
of hexadecimal and X (don't care) characters; otherwise, it is a binary
string made up of 0, 1, and X (don't care) characters.
Query Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA?
The :SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA? query returns the current
settings of the specified SPI data pattern resource in the binary string
format.
Return Format
<string><NL>
See Also "Introduction to :TRIGger Commands" on page 641
":SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh" on page 532
":SBUS<n>:SPI:SOURce:MOSI" on page 528
":SBUS<n>:SPI:SOURce:DATA" on page 525
NOTE
The :SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh should be set before
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA.
Vedere la pagina 530
1 2 ... 526 527 528 529 530 531 532 533 534 535 536 ... 969 970

Commenti su questo manuale

Nessun commento