Obsolete and Discontinued Commands 32
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 801
:MEASure:UPPer
(see page 860)
Command Syntax
:MEASure:UPPer <value>
The :MEASure:UPPer command sets the upper measurement threshold
value. This value and the LOWer value represent absolute values when the
thresholds are ABSolute and percentage when the thresholds are PERCent
as defined by the :MEASure:DEFine THResholds command.
Query Syntax
:MEASure:UPPer?
The :MEASure:UPPer? query returns the current upper threshold level.
Return Format
<value><NL>
<value> ::= the user-defined upper threshold in NR3 format
See Also "Introduction to :MEASure Commands" on page 330
":MEASure:LOWer" on page 791
":MEASure:THResholds" on page 794
NOTE
The :MEASure:UPPer command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:DEFine THResholds command (see page 337)
instead.