:DIGital<d> Commands 12
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 249
:DIGital<d>:DISPlay
(see page 860)
Command Syntax
:DIGital<d>:DISPlay <display>
<d> ::= 0 to (# digital channels - 1) in NR1 format
<display> ::= {{1 | ON} | {0 | OFF}}
The :DIGital<d>:DISPlay command turns digital display on or off for the
specified channel.
Query Syntax
:DIGital<d>:DISPlay?
The :DIGital<d>:DISPlay? query returns the current digital display setting
for the specified channel.
Return Format
<display><NL>
<display> ::= {0 | 1}
See Also "Introduction to :DIGital<d> Commands" on page 247
":POD<n>:DISPlay" on page 421
":CHANnel<n>:DISPlay" on page 226
":VIEW" on page 186
":BLANk" on page 162
":STATus" on page 183
NOTE
This command is only valid for the MSO models.