:SBUS<n> Commands 24
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 523
:SBUS<n>:SPI:FRAMing
(see page 860)
Command Syntax
:SBUS<n>:SPI:FRAMing <value>
<value> ::= {CHIPselect | {NCHipselect | NOTC} | TIMeout}
The :SBUS<n>:SPI:FRAMing command sets the SPI trigger framing value. If
TIMeout is selected, the timeout value is set by the
:SBUS<n>:SPI:CLOCk:TIMeout command.
Query Syntax
:SBUS<n>:SPI:FRAMing?
The :SBUS<n>:SPI:FRAMing? query returns the current SPI framing value.
Return Format
<value><NL>
<value> ::= {CHIP | NCH | TIM}
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:MODE" on page 648
":SBUS<n>:SPI:CLOCk:TIMeout" on page 522
":SBUS<n>:SPI:SOURce:FRAMe" on page 526
NOTE
The NOTC value is deprecated. It is the same as NCHipselect.