Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manuale di Servizio Pagina 778

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 970
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 777
778 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
32 Obsolete and Discontinued Commands
:CHANnel<n>:INPut
(see page 860)
Command Syntax
:CHANnel<n>:INPut <impedance>
<impedance> ::= {ONEMeg | FIFTy}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:INPut command selects the input impedance setting for
the specified channel. The legal values for this command are ONEMeg (1
M
) and FIFTy (50).
Query Syntax
:CHANnel<n>:INPut?
The :CHANnel<n>:INPut? query returns the current input impedance
setting for the specified channel.
Return Format
<impedance value><NL>
<impedance value> ::= {ONEM | FIFT}
NOTE
The :CHANnel<n>:INPut command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :CHANnel<n>:IMPedance command (see page 227)
instead.
Vedere la pagina 777
1 2 ... 773 774 775 776 777 778 779 780 781 782 783 ... 969 970

Commenti su questo manuale

Nessun commento