648 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
28 :TRIGger Commands
:TRIGger:MODE
(see page 860)
Command Syntax
:TRIGger:MODE <mode>
<mode> ::= {EDGE | GLITch | PATTern | TV | EBURst | RUNT | SHOLd
| TRANsition | SBUS{1 | 2} | USB}
The :TRIGger:MODE command selects the trigger mode (trigger type).
Query Syntax
:TRIGger:MODE?
The :TRIGger:MODE? query returns the current trigger mode. If the
:TIMebase:MODE is ROLL or XY, the query returns "NONE".
Return Format
<mode><NL>
<mode> ::= {EDGE | GLIT | PATT | TV | EBUR | RUNT | SHOL
| TRAN | SBUS{1 | 2} | USB}
See Also • "Introduction to :TRIGger Commands" on page 641
• ":TRIGger:SWEep" on page 650
• ":TIMebase:MODE" on page 631
Example Code
' TRIGGER_MODE - Set the trigger mode to EDGE.
myScope.WriteString ":TRIGger:MODE EDGE"
See complete example programs at: Chapter 37, “Programming Examples,”
starting on page 869