Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manuale di Servizio Pagina 522

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 970
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 521
522 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:SPI:CLOCk:TIMeout
(see page 860)
Command Syntax
:SBUS<n>:SPI:CLOCk:TIMeout <time_value>
<time_value> ::= time in seconds in NR3 format
The :SBUS<n>:SPI:CLOCk:TIMeout command sets the SPI signal clock
timeout resource in seconds from 100 ns to 10 s when the
:SBUS<n>:SPI:FRAMing command is set to TIMeout. The timer is used to
frame a signal by a clock timeout.
Query Syntax
:SBUS<n>:SPI:CLOCk:TIMeout?
The :SBUS<n>:SPI:CLOCk:TIMeout? query returns current SPI clock timeout
setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR3 format
See Also "Introduction to :TRIGger Commands" on page 641
":SBUS<n>:SPI:CLOCk:SLOPe" on page 521
":SBUS<n>:SPI:SOURce:CLOCk" on page 524
":SBUS<n>:SPI:FRAMing" on page 523
Vedere la pagina 521
1 2 ... 517 518 519 520 521 522 523 524 525 526 527 ... 969 970

Commenti su questo manuale

Nessun commento