Root (:) Commands 6
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 165
:MTEenable (Mask Test Event Enable Register)
(see page 860)
Command Syntax
:MTEenable <mask>
<mask> ::= 16-bit integer
The :MTEenable command sets a mask in the Mask Test Event Enable
register. Set any of the following bits to "1" to enable bit 9 in the
Operation Status Condition Register and potentially cause an SRQ (Service
Request interrupt to be generated.
Query Syntax :MTEenable?
The :MTEenable? query returns the current value contained in the Mask
Test Event Enable register as an integer number.
Return Format
<value><NL>
<value> ::= integer in NR1 format.
See Also "Introduction to Root (:) Commands" on page 154
":AER (Arm Event Register)" on page 156
Table 60 Mask Test Event Enable Register (MTEenable)
Bit Name Description When Set (1 = High = True), Enables:
15-11 --- --- (Not used.)
10 Auto
Mask
Auto Mask Created Auto mask creation completed.
9 --- --- (Not used.)
8 Started Mask Testing Started Mask testing started.
7-2 --- --- (Not used.)
1 Fail Mask Test Fail Mask test failed.
0Comple
te
Mask Test Complete Mask test is complete.
To MTE bit in
Operation Status
Condition Register
+
OR
:MTEenable
:MTEenable?
Mask Test Event Enable (Mask) Register
1023456798101112131415
:MTERegister[:EVENt]?
Mask Test Event Event Register
Fail
Com-
plete
Auto
Mask
Started
Vedere la pagina 164