:SAVE Commands 23
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 439
:SAVE:IMAGe:FORMat
(see page 860)
Command Syntax
:SAVE:IMAGe:FORMat <format>
<format> ::= {{BMP | BMP24bit} | BMP8bit | PNG}
The :SAVE:IMAGe:FORMat command sets the image format type.
Query Syntax
:SAVE:IMAGe:FORMat?
The :SAVE:IMAGe:FORMat? query returns the selected image format type.
Return Format
<format><NL>
<format> ::= {BMP | BMP8 | PNG | NONE}
When NONE is returned, it indicates that a waveform data file format is
currently selected.
See Also "Introduction to :SAVE Commands" on page 434
":SAVE:IMAGe[:STARt]" on page 437
":SAVE:IMAGe:FACTors" on page 438
":SAVE:IMAGe:INKSaver" on page 440
":SAVE:IMAGe:PALette" on page 441